6.3.8 High Pressure
The
High pressure
connection can be used to measure
pressures greater than 150 mbar.
If you prefer for this connection a DISS O
2
Connection, there is
an appropriate adapter available.
Range:
0...10 bar
Accuracy:
± 1% of reading or ± 10 mbar
High pressure
When measuring below 150 mbar it is
recommended to use the Differential Pressure port
± 150 mbar since the accuracy is up to 100 times higher.
Pressures higher than 15 bar will destroy the sensor!