U
SER
M
ANUAL
F
LOW
A
NALYZER
TM
Page 21 / 84
6.3.7 Pressure sensor ± 1
bar
(PFC-3000V)
There is an additional pressure sensor ± 1 bar for the
PFC-
3000V
. The sensor is connected to the specified connector and
is marked with a red ring.
Range:
± 1000 mbar
Accuracy:
± 0.5% of reading or ± 2 mbar
Pressure sensor ±1 bar
When using one of the pressure options (low P or ±
1bar), the sensor (± 150 mbar) will be connected to the
remaining connector where as the second port of the sensor
is measuring against ambient. The measurement range
remains the same.