BC Biomedical DPM-2001 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

DIGITAL PRESSURE METERS 

 

 

 

 

 

 

 

DPM-2001

 

 

 

 

 

 

 

 

  DPM-2001PLUS                 DPM-2100 

 

USER MANUAL

 

Содержание DPM-2001

Страница 1: ...DIGITAL PRESSURE METERS DPM 2001 DPM 2001PLUS DPM 2100 USER MANUAL...

Страница 2: ...0 SETUP MODE 12 INPUTS 13 STATUS DISPLAYS 14 COMMUNICATIONS 15 MANUAL REVISIONS 20 WARRANTY 20 SPECIFICATIONS 21 DRAWINGS DPM 2001 PHYSICAL DIAGRAM 23 DPM 2001 PLUS PHYSICAL DIAGRAM 24 DPM 2100 PHYSIC...

Страница 3: ...IONAL INC RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFO...

Страница 4: ...OTS TO TURN 5 DIGIT LCD PLUS SCALE INDICATION BATTERY LIFE DISPLAY 0 to 100 PROGRAMMABLE DIGITAL FILTER LCD CONTRAST IS SOFTWARE ADJUSTABLE 16 BIT PRESSURE MEASUREMENT DIGITAL ZERO ADJUST DPM 2001 PLU...

Страница 5: ...4 OPTIONAL ACCESSORIES BE2000PU 120 VAC BATTERY ELIMINATOR U S BE2000PE 220 VAC BATTERY ELIMINATOR EUROPE DPM2000C RS 232 CABLE BC20 01004 HARD CARRYING CASE BC20 01005 UNIVERSAL PRESSURE ADAPTER KIT...

Страница 6: ...rated for 20 degrees Celsius TEMPERATURE SCALE MODEL DPM 2100 ONLY The temperature can be displayed in either degrees Celsius or Fahrenheit When temperature is displayed the RANGE key will toggle the...

Страница 7: ...ssure is being displayed the Setup Mode will be entered SETUP MODE Pressing this key while in the Setup Mode will sequence the display through the available parameters OFF RESET The function of this k...

Страница 8: ...erating Mode as follows PRESSURE MEASUREMENT If this key is pressed while the pressure is being displayed the unit will display the percent 0 to 100 of battery life remaining NOTE If the optional wall...

Страница 9: ...he Setup Mode will sequence the display through the available parameters OFF RESET The function of this key is dependant on the Current Operating Mode as follows PRESSURE MEASUREMENT If this key is pr...

Страница 10: ...d the unit will display the minimum pressure detected since the capture register was last reset SETUP MODE If this key is pressed in the Setup Mode the value of the displayed setting will decrement Pr...

Страница 11: ...etup Mode will sequence the display through the available parameters OFF RESET The function of this key is dependant on the Current Operating Mode as follows PRESSURE MEASUREMENT If this key is presse...

Страница 12: ...ressure is being displayed the unit will display the minimum pressure detected since the capture register was last reset TEMPERATURE MEASUREMENT If this key is pressed while the temperature is being d...

Страница 13: ...mines the period of inactivity before the meter is turned OFF A timer is started when the meter is turned ON and is reset each time a key is pressed When the timer reaches the value set in this parame...

Страница 14: ...VDC Battery Eliminator power supply that may be used for continuous run applications It bypasses the internal battery when plugged in RS232 This is the serial interface connector MODELS 2001 PLUS AND...

Страница 15: ...level LOW BAT is displayed for two seconds once a minute until the battery is replaced DEF CAL This status screen displays on Power Up if the unit is out of Calibration The unit will load default val...

Страница 16: ...o parity and no handshaking The following section describes the JPC Protocol used by the meter The JPC Protocol consists of 6 basic commands R READ W WRITE U UPLOAD Q QUICKSEND V VERSION X CANCEL The...

Страница 17: ...icates that the command is to be activated The WRITE command allows the user to update the system settings The write command is entered as a letter followed by 7 numbers followed by a carriage return...

Страница 18: ...line feed 0A Data Sent Data Returned Meaning R03 cr R03 cr lf Echo of Command Sent 00010 cr lf Auto Off is set to 10 minutes W0300000 cr W0300000 cr lf Echo of Command Sent Change Auto off to disable...

Страница 19: ...and section for details on the data structure QUICKSEND COMMAND Quicksend is a feature that allows the user to receive an automatic update of all of the meter data without any user interaction When th...

Страница 20: ...ransmitted For the most part an incorrect command will simply be ignored and the meter will prepare for an additional command However a command may be cancelled midstream by transmitting an X ASCII Th...

Страница 21: ...D THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANT...

Страница 22: ...CCURACY 0 1 FS TEMPERATURE RANGE 0 0 to 100 0 C DPM 2100 MODELS ONLY 32 0 to 212 0 F TEMPERATURE RESOLUTION 0 1 C 0 1 F TEMPERATURE ACCURACY 0 5 FS DISPLAY Six 0 35 Inch High Seven Segment LCD Uniplan...

Страница 23: ...Temperature Probes or equivalent PRESSURE MEDIA Any pure fluid or gas that is compatible with Pyrex Glass Silicon Alumina Ceramic Epoxy RTV gold aluminum and nickel POWER LINE 9VDC Center Negative BAT...

Страница 24: ...23...

Страница 25: ...24...

Страница 26: ...25...

Страница 27: ...26 NOTES...

Отзывы: