OTH-SEM250-0356 Rev D
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Recommended separation distance:
P
d
⋅
=
2
.
1
150kHz to 80 MHz
P
d
⋅
=
2
.
1
80 MHz to 800 MHz
P
d
⋅
=
3
.
2
800MHz to 2.7 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitter, as
determined by an electromagnetic site survey a,
should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity of
equipment marked with the following symbol: