DMR DEX 3.3 V-5 25-May-2022 (DEX-14)

 

 

 

Instructions for Use 

Electrophysiology Cable (DEX-14)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

English 

Carefully read all instructions prior to use.  Observe all contraindications, warnings 

and precautions noted in these instructions.  Failure to do so may result in patient 

Complications. 

Caution:

 Federal (U.S.A.) law restricts this device to sale by or on the order of a 

physician. 

I. 

DEVICE DESCRIPTION  

The Electrophysiology Cable (DEX-14) connects the Baylis Medical Company Inc. 

EPstarFixed  Electrophysiology Catheter with Lumen (DLF) to diagnostic 

electrophysiology equipment (diagnostic EP equipment), such as an 

electrocardiography system and/or cardiac stimulator. This cable conducts 

intracardiac potentials between the DLF and a catheter input module of the 

diagnostic EP equipment. 

Detailed information concerning the DLF is contained in a separate manual that 

accompanies the DLF Instructions for Use. 

The dimensions for the Electrophysiology Cable (DEX-14) can be found on the 

device label and in section VII  “PRODUCT SPECIFICATIONS”. The 

Electrophysiology Cable (DEX-14) has a single 14-pin connector on one end that 

mates with the DLF, and fourteen single-pin connectors on the other end that mate 

with the catheter input module of diagnostic EP equipment. 

II. 

INDICATIONS FOR USE  

The  Electrophysiology Cable  (DEX-14)  used with the EPstar  Fixed 

Electrophysiology Catheter with Lumen can be used in the evaluation of a variety 

of cardiac arrhythmias from endocardial and intravascular sites. 

III. 

CONTRAINDICATIONS   

The Electrophysiology Cable (DEX-14) is not recommended for use with any other 

electrophysiology catheter. 

IV. 

WARNINGS  

 

The Electrophysiology Cable (DEX-14) is a single-use device. Do not attempt 

to  re-use or re-sterilize the device.  Re-sterilization of the cable may 

compromise the device performance. 

 

The  Electrophysiology Cable  (DEX-14)  must only be used with a BMC 

EPstar Fixed Electrophysiology Catheter with Lumen (DLF). Attempts to use 

it with other diagnostic catheters and devices can result in complications. 

 

Laboratory staff and patients can undergo significant X-ray exposure during 

diagnostic catheterization procedures due to the continuous usage of 

fluoroscopic imaging.  This exposure can result in acute radiation injury as 

well as increased risk for somatic and genetic effects.  Therefore, adequate 

measures must be taken to minimize this exposure. 

 

Avoid exposing the connectors of the Electrophysiology Cable (DEX-14) to 

any fluids during a procedure. Exposure to fluids may result in shorting of 

electrical signals. 

 

Diagnostic electrophysiology equipment is susceptible to electromagnetic 

interference. Do not operate near equipment that generate strong 

electromagnetic fields. 

 

Do not alter this device in any way. 

V. 

PRECAUTIONS  

 

Do not attempt to use the Electrophysiology Cable  (DEX-14)  or ancillary 

equipment before thoroughly reading the accompanying Instructions for Use. 

 

Use only for cardiac electrophysiological examinations. 

 

Catheterization procedures should be performed only by trained physicians 

in a fully equipped catheterization laboratory. 

 

The sterile packaging should be visually inspected prior to use to detect any 

compromise.  Ensure that the packaging has not been damaged.  Do not use 

the equipment if the packaging has been compromised. 

 

Visually inspect the cable to ensure there is no damage to the insulating 

material.  Do not use the cable if there is any damage. 

 

Never disconnect the Electrophysiology Cable (DEX-14) from the catheter 

by pulling on the cable.  Failure to disconnect the cable properly may result 

in damage to the cable. 

 

Do not twist the Electrophysiology Cable (DEX-14) while inserting it to or 

disconnecting it from the catheter connector.  Twisting the cable may result 

in damage to the pin connectors. 

 

Do not bend the cable excessively.  Excessive bending or kinking of the 

cable may damage the integrity of the cable and may lead to patient injury.  

Care must be taken when handling the cable. 

 

Adequate filtering must be used to allow continuous monitoring of the 

electrocardiogram (ECG) signals during the procedure. 

 

Store under stable conditions, avoiding vibration and shock (including during 

transportation). 

 

Avoid exposure to direct sunlight.  

 

Use only with legally marketed diagnostic EP equipment. 

Baylis Medical Company Inc. relies on the physician to determine, assess and 

communicate to each individual patient all foreseeable risks of an 

electrophysiology procedure. 

VI. 

ADVERSE EVENTS  

There are no adverse events associated with the use of this device. Adverse 

events associated with the use of this device as part of a larger system are 

indicated for the EPstar Fixed Electrophysiology Catheter with Lumen (DLF). 

VII. 

PRODUCT SPECIFICATIONS 

Model Number 

DEX-14 

Overall Useable Length 

7.5 feet (2.3m) 

Catheter Connector 

14-pin (Plug) 

Diagnostic Equipment Connectors 

Single-pin (Plug, DIN 42802-2) × 14 

Compatible Catheter Models 

DLF-6-10-55-95R 

VIII. 

INSPECTION PRIOR TO USE 

Perform the following checks before the patient is presented for the procedure. 

These tests will allow you to verify that the equipment you will use is in proper 

working order.  Do these tests in a sterile environment.  Do not use defective 

equipment. 

KEY ITEMS 

QUESTION? 

WARNINGS AND EXPLANATIONS 

Sterility 

Is the 

connector 

cable sterile? 

The  Electrophysiology Cable  (DEX-14)  is 

supplied sterile for its initial use.  Inspect the 

packaging to ensure the package has not been 

damaged and sterility has not been compromised. 

Do not use if cable is not sterile.  

Visual Check 

Have you done 

a visual check 

on the entire 

system? 

Ensure connectors and the cable have no visible 

damage, such as discoloration, cracks, label 

fading, cable splice, or kinks. Do not use 

damaged equipment. 

IX. 

EQUIPMENT REQUIRED 

Diagnostic electrophysiology procedures should be performed in a specialized 

clinical setting equipped with a fluoroscopy unit, radiographic table, physiologic 

recorder, emergency equipment and instrumentation for gaining vascular access. 

X. 

DIRECTIONS FOR USE 

All instructions for equipment required should be carefully read, understood, and 

followed.  Failure to do so may result in complications.  

Ensure that the EPstar  catheter is properly positioned in the patient and the 

diagnostic EP equipment is set up before connecting the Electrophysiology Cable 

(DEX-14). 

1. 

Connect the 14-pin catheter connector of the cable to the connector of the 

EPstar  catheter. The catheter  connector is circular and keyed for proper 

alignment. Line up the arrow markings on the connector with the catheter 

receptacle and gently push in until the connector are mated. Any attempt to 

connect the cable otherwise will damage the pins on the connector. Use of 

excessive force may result in damage to the connector pins. 

2. 

Connect the single-pin connectors at the diagnostic EP equipment end of the 

cable to the catheter input module of the diagnostic EP equipment. Observe 

the marking on each single-pin connector as they correspond to the 

electrode position of the catheter (“D” is associated with the most Distal 

electrode, where the remaining numbers indicate the relative position of the 

electrode from the tip of the catheter). 

 

Baylis Medical Company Inc. 

5959 Trans-Canada Highway 

Montreal, Quebec, Canada, H4T 1A1 

Tel: (514) 488-9801/ (800) 850-9801 Fax: (514) 488-7209 

www.baylismedical.com

 

© Copyright Baylis Medical Company Inc., 2022 

 

The Baylis Medical logo is a trademark and/or a registered trademark of Baylis 

Medical Company Inc. in the United States of America and/or other countries. 

All other trademarks or registered trademarks are property of their respective 

owners.

 