178
Rev. C
Additional Wash
Required
A change in transmittance was detected by the optical clarity test
that occurs during the wash after a CO-ox sample analysis. The
system repeated the wash to attempt to correct the error. If the
repeated wash fails the clarity test again, the system displays a
D70 Optical Error: 2 message. Refer to the solutions for a D70
Optical Error on page 205.
Service Solutions
Refer to service solutions for D70 Optical Error: 2.
Analysis is turned
off by a remote
computer.
The Rapidlink data management system or the LIS connected to
the Rapidpoint 400 series
system has turned analysis functions
off for the Rapidpoint 400 series
system. You can perform other
tasks, such as recalling results, replacing cartridges, and
accessing the Setup options. Contact your system supervisor for
assistance if you need to analyze samples. The Analysis Turned
Off message appears in the events log.
Bubbles in
the Sample. The
system cannot
complete analysis.
Touch Continue
to begin the
sequence to
clear the system.
Replace the
sample port when
prompted.
The system cannot analyze the current patient or QC sample
because it detects bubbles in the sample. This can occur if the
sample contained bubbles when you introduced it into the
sample port or if an obstruction is present. The Bubbles in
Sample message appears in the events log.
Introducing the jagged end of capillary tubes into the sample port
can also cause this problem. The sharp edges of capillary tubes
can damage the sample port, affecting sample aspiration.
1. Touch
Continue
to remove the message, then replace the
sample port when prompted. Refer to
Replacing the Sample
Port
in
Replacing Components
.
2. If prompted, replace the cartridges as described in
Replacing
the Measurement and Wash/Waste Cartridges
in Section 3 in
the
Rapidpoint 400 Series Operator’s Manual
.
3. To avoid bubbles in patient samples, ensure that you use the
recommended collection, storage, handling, and mixing
techniques as described in Section 1,
Collecting Patient
Samples
in the
Rapidpoint 400 Series Operator’s Manual
.
For QC samples, ensure that you use the storage and handling
techniques recommended by the manufacturer.
(Continued)
Message
Probable Cause and Corrective Action
Содержание RapidPoint 400 Series
Страница 1: ...Service Manual 105953 Rev C 2003 02...
Страница 29: ...Rev C Mechanical Descriptions 29 Rapidpoint 400 Series System left side 1 Diskette drive...
Страница 62: ...62 Rev C...
Страница 112: ...112 Rev C...
Страница 114: ...114 Rev C...
Страница 116: ...116 Rev C Rapidpoint 400 Series Systems Menu Map...
Страница 117: ...Rev C System Diagrams and Cable Drawings 117 Fluidic Diagrams...
Страница 118: ...118 Rev C...
Страница 119: ...Rev C System Diagrams and Cable Drawings 119...
Страница 120: ...120 Rev C...
Страница 121: ...Rev C System Diagrams and Cable Drawings 121...
Страница 122: ...122 Rev C...
Страница 123: ...Rev C System Diagrams and Cable Drawings 123...
Страница 124: ...124 Rev C...
Страница 125: ...Rev C System Diagrams and Cable Drawings 125...
Страница 126: ...126 Rev C...
Страница 127: ...Rev C System Diagrams and Cable Drawings 127...
Страница 128: ...128 Rev C Rapidpoint 400 Series Systems Electronic Block Text Diagram...
Страница 129: ...Rev C System Diagrams and Cable Drawings 129 Rapidpoint 400 Series Systems Cable Interconnect Diagram...
Страница 130: ...130 Rev C Cable Drawings A C Power Harness Assembly 570250 Cable Harness Outboard Assembly 113509...
Страница 131: ...Rev C System Diagrams and Cable Drawings 131 D C Power Harness Assembly 113836...
Страница 132: ...132 Rev C Diskette Drive Cable Assembly 570255 Diskette Power Harness 570254...
Страница 133: ...Rev C System Diagrams and Cable Drawings 133 Hard Drive Cable Assembly 570257 Inboard 1 Harness 113770...
Страница 134: ...134 Rev C Inboard 2 Harness 113523 LCD Backlight Cable Assembly 108771...
Страница 135: ...Rev C System Diagrams and Cable Drawings 135 Preamp Module Cable Assembly 570269...
Страница 136: ...136 Rev C UIP Display Cable Assembly 570263 UIP LPT1 COM2 Cable Assembly 570267...
Страница 137: ...Rev C System Diagrams and Cable Drawings 137 UIP Utility Cable Assembly 570268 UIP COM1 Inboard Cable Assembly 570272...
Страница 138: ...138 Rev C Flex Circuit Display 570516 100 Pin Cable Assembly 111326 Patch Cord Cable 122254...
Страница 139: ...Rev C System Diagrams and Cable Drawings 139 Fiber Bundle Assembly 1315271...
Страница 140: ...140 Rev C...
Страница 142: ...142 Rev C...
Страница 148: ...148 Rev C...
Страница 154: ...154 Rev C...
Страница 258: ...258 Rev C...
Страница 260: ...260 Rev C...
Страница 262: ...262 Rev C Handle Pawl...
Страница 263: ...Rev C Illustrated Parts Lists 263 Wash Waste Switch Rubber Feet...
Страница 264: ...264 Rev C Display Display Bracket...
Страница 267: ...Rev C Illustrated Parts Lists 267 Front Door Door Hook...
Страница 268: ...268 Rev C Cartridge Interface Assembly Front Door Switch...
Страница 269: ...Rev C Illustrated Parts Lists 269 ESD Ground Strap Polychromator Module 1 CO ox polychromator module...
Страница 270: ...270 Rev C Diskette Drive Diskette Drive Side 1 Diskette drive cable 2 Diskette drive fascia 3 Diskette drive...
Страница 271: ...Rev C Illustrated Parts Lists 271 Diskette Drive Diskette Drive Cable Shield...
Страница 273: ...Rev C Illustrated Parts Lists 273 Connector UIP PCB UIP PCB with Cables...
Страница 274: ...274 Rev C Main PCB Main PCB Attached to Frame...
Страница 275: ...Rev C Illustrated Parts Lists 275 AQC Frame Assembly AQC Frame Assembly 1 AQC frame assembly 2 Main Interconnect board...
Страница 276: ...276 Rev C AQC Latch Assembly Front View AQC Latch Assembly Rear View 1 AQC latch assembly...
Страница 277: ...Rev C Illustrated Parts Lists 277 Cover...
Страница 282: ...282 Rev C Pump Housing Peri pump Motor and Mount...
Страница 283: ...Rev C Illustrated Parts Lists 283 Peri pumps Spring Plate...
Страница 284: ...284 Rev C Connector Block Cartridge Handle...
Страница 285: ...Rev C Illustrated Parts Lists 285 Preamp Cable Cover...
Страница 289: ...Rev C Illustrated Parts Lists 289 Power Module Frame Power Entry with Switch...
Страница 290: ...290 Rev C Rear Interconnect Board Fan...
Страница 291: ...Rev C Illustrated Parts Lists 291 Power Supply Patch Cord with Ferrite Filter...
Страница 292: ...292 Rev C Hard Drive...
Страница 294: ...294 Rev C CO ox Module Components Illustrations Sample Chamber Interface Assembly RSL Part 131530...
Страница 296: ...296 Rev C Neon Board Assembly Part Number 131526...
Страница 298: ...298 Rev C...
Страница 302: ...302 Rev C...
Страница 351: ...Rev C Replacing Components 351 a Remove the four thread forming screws from the motor mount...
Страница 380: ...380 Rev C...
Страница 381: ...Rev C Spare Parts 381 Spare Parts Spare Parts Listing 383...
Страница 382: ...382 Rev C...
Страница 386: ...386 Rev C...
Страница 400: ...400 Rev C...
Страница 426: ...426 Rev C...
Страница 432: ...432 Rev C...
Страница 438: ...438 Rev C...
Страница 458: ...458 Rev C...