MEDRAD
®
Stellant CT Injection System with Certegra
®
Workstation
9 - 36
Figure 9 - 2: Create a Protocol
i.
Select an arrow to create a phase.
(1)
ii.
Choose a phase type from the list.
(2)
iii.
Add/edit volumes and flow rates.
(3)
iv.
Select a pressure limit.
(4)
v.
Press Lock to lock the protocol.
(5)
vi.
Refer to
for further information.
2.
Select
Lock
on the display.
NOTE:
The icon on the display changes to
Arm
, and the injector head
Protocol Lock
indicator
illuminates.
9.2 Scan Room Preparation
4
5
2
1
3
Air Embolism Hazard - Serious patient injury or death may result.
•
Expel all trapped air from the syringe(s), connectors, tubing, and catheter before
connecting the system to the patient. Carefully read the instructions for loading and the
use of the MEDRAD
®
FluiDots (where applicable) to reduce the chance of air embolism.
•
The presence of rounded FluiDots does not indicate the total absence of air bubbles in the
syringe tip. FluiDots must be viewed in a properly illuminated environment with a light
source behind the operator providing enough light to permit easy viewing.
•
To minimize air embolization risks, ensure that one operator is designated the
responsibility of filling the syringe(s). Do not change operators during the procedure. If an
operator change must occur, ensure that the new operator verifies that the fluid path is
purged of air.
•
To minimize the possibility of inadvertent aspiration and injection, ensure the patient is
disconnected from the injector when utilizing the forward and reverse piston controls.
Biological Contamination Hazard - Serious patient and/or worker injury or death may result.
•
Properly discard fluid source containers and disposable items after use (refer to
disposable label for specifics), or if there is any possibility that contamination may have
occurred.
•
Do not reuse single-use fluid sources. Refer to the fluid source manufacturers’
instructions for use.
WARNING
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......