Page 1 of 2 

Urgent Field Safety Notice 

 

PrisMax, V2, ROW  
FA-2020-064  
Safety Alert 

 

January 2021  

Dear Healthcare Provider 

Problem 
Description 

 

In an effort to keep you informed of important product information that can help ensure 
the safe and effective use of our products, Baxter Healthcare is issuing a Safety Alert 
for the PrisMax System. Baxter has received reports of users who have confused the 
acronyms for Pre-Blood Pump (PBP) and Patient Fluid Removal (PFR) when entering 
the prescriptions on the PrisMax Graphical User Interface (GUI).  
 
During setup, the user needs to enter the Pre-Blood Pump (PBP) and the Patient Fluid 
Removal (PFR) flow rates. 

Please ensure the appropriate values are entered in the 

respective fields.

 Color coding and an additional help button are available for further 

support when entering the prescription. Please confirm the entered Pre-Blood Pump 
and  Patient  Fluid  Removal  values  on  the  Review  screen.  Switching  these  two  flow 
rates may result in excessive fluid removal from the patient.  
 
The acronyms for PBP and PFR are defined 

in the Operator’s Manual as well as during 

therapy on the GUI screen. Refer to the enclosure for details. 
 

To ensure patient safety, the device should only be used by a trained operator. 

 
 

Affected 
Product  

 

Product Code 

Product Description 

Serial Numbers 

955558 

PrisMax V2, ROW 

All 

 
Hazard 
Involved 

 
Incorrect  therapy  settings  could  lead  to  unintended  excessive  fluid  removal  during 
treatment. This has the potential to lead to hypotension. There have been two reports 
of serious injury related to this issue. 
 

Actions to be 
Taken by 
Customers 

 

 
 
 
 
 
 
 

 

 
 
 

 

1.  To ensure patient safety, the device should only be used by a trained operator per 

the instructions in the Operator’s Manual. 

Please ensure that every operator of 

this device is made aware of this Safety Alert.

 

 
2. 

If you purchased this product directly from Baxter, complete the enclosed 
Baxter Customer Reply Form and return it to Baxter by faxing it or scanning 
it.

  Returning  the  customer  reply  form  promptly  will  confirm  your  receipt  of  this 

notification and prevent you from receiving repeat notices. 

 
3.  If  you  purchased  this  product  from  a  distributor,  please  note  that  the  Baxter 

customer reply form is not applicable. If a reply form is provided by your distributor 
or wholesaler, please return it to the supplier according to their instructions. 

 

4.  If you distribute this product to other facilities or departments within your institution, 

please forward a copy of this communication to them. 

 