Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 120
53007410 Rev. C
Password Expiration
Select
Use password expiration
to place an expiration date on each password, if your facility’s
protocol requires that user passwords must be changed on a regular basis.
NOTE:
By default Use password expiration checkbox is checked.
For
Days password valid
, enter the number of days after which the password expires.
Tip!
Baxter recommends entering 90 (or the number specified by your facility’s protocol) for the
Days password valid
.
Order Entry Serial Number
For
Serial number length
, enter the number of maximum length of serial number for a formula
file.
NOTE:
If the Order Entry Serial Number is more than 36 characters, part of it may be truncated
in the reports
.
Barcode Reader
Select
Enable barcode reader
to allow the use of a barcode reader for scanning labels during
verification of the setup and for loading formulas. If you select this check box, these options
become available:
•
Select
Use barcode verification
if you want to require the use of a barcode reader for
scanning labels during verification of the setup.
•
Select
Require barcode to initiate compounding
if you want to require the use of a
barcode reader for loading formulas. If this box is selected, the user cannot manually
select a saved formula.
WARNING
It is important to use a barcode reader for scanning labels during verification of the
setup and for loading formulas.
NOTE:
Compounder supports both 1D & 2D GS1 barcode formats. However to enable 2D
barcode reading a 2D barcode reader (part number # 6500-0500) is needed.
Default User
If your facility does not require users to log in, select a
Default User
who is logged in
automatically at startup.
Tip!
Baxter does not recommend using this feature.
Cosignature
Select
Required for Configuration Verification
if you want to require a second user to log in
and verify the configuration. If this option is not selected, the same user can set up and verify
the configuration.
Tip!
Baxter strongly recommends requiring a cosignature.
Содержание ExactaMix 2400 Compounder
Страница 2: ......
Страница 170: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 162 53007410 Rev C Sample MixCheck Report...
Страница 177: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 169 Sample Formula Report...
Страница 180: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 172 53007410 Rev C Sample Log Report...
Страница 185: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 177 Sample Formulary Report...
Страница 199: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 191 Sample Bag Usage Report...