EASYSPRAY 

PRESSURE REGULATOR

 

QUICK REFERENCE GUIDE 

For Tech Support Call 1-888-229-0001,option 4

Selected Important Risk Information 

Air or gas embolism has occurred when fibrin sealant was administered using pressurized gas.  This can occur  
if a spray device is used at higher than recommended pressures and in closer than recommended proximity to the tissue surface.

TIPS

 

The EASYSPRAY device will continue to emit gas for a brief period after the thumb is removed from the clip/plunger.  
This delay helps to avoid clogging of the spray head.

Instructions for Circulating Nurse | 

EASYSPRAY Device

Instructions for Scrub Nurse | 

Spray Set for TISSEEL [Fibrin Sealant]

Pass the assembled TISSEEL syringe to 
the surgeon for spray application.

Pass the end of the connection tube with 
the sterile filters to the circulating nurse.

6

Attach the clip (on the end of the sensor 
line) by sliding it into the grooves located 
on the top of the syringe plunger.

Fit the connection tube of the spray set to 
the Luer-lock connector on the underside of 
the spray head.

4

Fasten the pull strap to the double 
syringe system to assure the spray head 
is tightly secured.

3

Firmly attach the spray head to the 
nozzle of the syringes by aligning the 
blue dot on the spray head to the blue 
dot on the syringes.

2

Prepare TISSEEL [Fibrin Sealant] according 
to the instructions in the package insert.

1

Instructions for Surgeon

A spraying distance of 10 –15 cm is 
recommended. Note that the spraying 
distance should not be less than 10 cm 
(3.9”) and the maximum pressure should 
not exceed 2 bars (29.5psi).  

To activate the flow of gas, occlude the 
opening in the clip center  with thumb. 
To begin application, gently depress the 
TISSEEL syringe plunger.

2

Please see Indications and Detailed Important Risk Information for TISSEEL [Fibrin Sealant] on the reverse side.

Rx Only

:  For safe and proper use of these devices, please refer to the appropriate full device Instructions for Use included.

1

Insert 9V battery into the  EASYSPRAY 
pressure regulator device. 

Check the gauge on the EASYSPRAY
device for the appropriate pressure
range of 1.5-2.0 bars (21.8-29 psi).
Adjust pressure setting by turning the
black pressure control knob.

6

Turn the on/off switch on the front side 
of the EASYSPRAY device to the ON posi-
tion. The low-battery indicator light will 
only illuminate when the battery is weak 
and when the gas flow is activated. If 
the battery is completely dead, the low-
battery indicator will not illuminate.

5

5

Connect Spray Set filters to  EASYSPRAY 
device. Connect the blue filter to the blue 
female Luer connector and the clear filter 
 to the male Luer connector. 

4

Connect the adapter attached to the 
black tubing, located at the side of the 
EASYSPRAY device, to an appropriate gas 
source. Pressurized gas supply must be 
set between 3.5 and 7 bar (51-100 psi).

3

Connect EASYSPRAY device to IV pole or 
cart rail using the clamps on the back of 
the device.

2

1

Do Not connect to an oxygen source.
Do Not use rechargable batteries.

For Topical Use Only

For Topical Use Only

For Topical Use Only

For Topical Use Only

For Topical Use Only

For Topical Use Only