EASYSPRAY
PRESSURE REGULATOR
QUICK REFERENCE GUIDE
For Tech Support Call 1-888-229-0001,option 4
Selected Important Risk Information
Air or gas embolism has occurred when fibrin sealant was administered using pressurized gas. This can occur
if a spray device is used at higher than recommended pressures and in closer than recommended proximity to the tissue surface.
TIPS
The EASYSPRAY device will continue to emit gas for a brief period after the thumb is removed from the clip/plunger.
This delay helps to avoid clogging of the spray head.
Instructions for Circulating Nurse |
EASYSPRAY Device
Instructions for Scrub Nurse |
Spray Set for TISSEEL [Fibrin Sealant]
Pass the assembled TISSEEL syringe to
the surgeon for spray application.
Pass the end of the connection tube with
the sterile filters to the circulating nurse.
6
Attach the clip (on the end of the sensor
line) by sliding it into the grooves located
on the top of the syringe plunger.
Fit the connection tube of the spray set to
the Luer-lock connector on the underside of
the spray head.
4
Fasten the pull strap to the double
syringe system to assure the spray head
is tightly secured.
3
Firmly attach the spray head to the
nozzle of the syringes by aligning the
blue dot on the spray head to the blue
dot on the syringes.
2
Prepare TISSEEL [Fibrin Sealant] according
to the instructions in the package insert.
1
Instructions for Surgeon
A spraying distance of 10 –15 cm is
recommended. Note that the spraying
distance should not be less than 10 cm
(3.9”) and the maximum pressure should
not exceed 2 bars (29.5psi).
To activate the flow of gas, occlude the
opening in the clip center with thumb.
To begin application, gently depress the
TISSEEL syringe plunger.
2
Please see Indications and Detailed Important Risk Information for TISSEEL [Fibrin Sealant] on the reverse side.
Rx Only
: For safe and proper use of these devices, please refer to the appropriate full device Instructions for Use included.
1
Insert 9V battery into the EASYSPRAY
pressure regulator device.
Check the gauge on the EASYSPRAY
device for the appropriate pressure
range of 1.5-2.0 bars (21.8-29 psi).
Adjust pressure setting by turning the
black pressure control knob.
6
Turn the on/off switch on the front side
of the EASYSPRAY device to the ON posi-
tion. The low-battery indicator light will
only illuminate when the battery is weak
and when the gas flow is activated. If
the battery is completely dead, the low-
battery indicator will not illuminate.
5
5
Connect Spray Set filters to EASYSPRAY
device. Connect the blue filter to the blue
female Luer connector and the clear filter
to the male Luer connector.
4
Connect the adapter attached to the
black tubing, located at the side of the
EASYSPRAY device, to an appropriate gas
source. Pressurized gas supply must be
set between 3.5 and 7 bar (51-100 psi).
3
Connect EASYSPRAY device to IV pole or
cart rail using the clamps on the back of
the device.
2
1
Do Not connect to an oxygen source.
Do Not use rechargable batteries.
For Topical Use Only
For Topical Use Only
For Topical Use Only
For Topical Use Only
For Topical Use Only
For Topical Use Only