Cause:
A closed clamp, obstruction, or kink in the administration set is preventing
fluid flow between the source container and the pump.
Action:
Correct the problem causing the occlusion.
• Ensure the complete insertion of the spike into the source container.
• Inspect the administration set above the pump for closed clamps or kinks.
• Ensure that BURETROL administration sets or source containers are vented.
After correcting the problem, press the appropriate
Channel Select
key (for a
triple channel pump), press the
Primary
or
Secondary
soft key, then press
START
to resume the infusion.
Troubleshooting Alerts
The alert tone can be silenced for two minutes by pressing the
Alarm
Silence
key.
Advance Air
Pump module display message:
ADV AIR
Cause:
The pump’s Advance Air feature is being used to move an air bubble
through the tubing.
Action:
The
Advance Air
alert is active when the
Advance Air
soft key is
pressed. This alert clears when the
Done
soft key is pressed.
Channel Stopped
Pump module display message:
STOPPED
Cause:
The pump is ON and the infusion is not running.
Action:
Complete the remaining programming steps and press the
START
key, or
power off the pump.
LOW Battery - Plug in Now
Pump module display message
: the current infusion information
Cause:
The charge remaining in the batteries has 30 minutes of infusion time left.
The time remaining shown on the Main Display decrements if the pump is not
plugged in.
Action:
Plug the pump into an AC power source as soon as possible.
KVO: Volume Remaining = 0
Pump module display message:
KVO=x.x/label
(where x.x = the KVO rate)
Cause:
The Volume to be infused has reached zero and the pump is infusing at the
KVO rate (or programmed rate, whichever is lower).
Action:
Do one of the following as appropriate:
• Prepare a new infusion.
• For triple channel pumps, stop the channel and place it in Standby mode.
• Power off the pump.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...