![baxter Colleague 3 CXE Скачать руководство пользователя страница 215](http://html1.mh-extra.com/html/baxter/colleague-3-cxe/colleague-3-cxe_operators-manual_2712231215.webp)
071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-19
Quick Reference Guide
Unloading the Administration Set
Reset the Manual Tube Release as follows:
1.
Close the regulating clamp on the administration set.
Ensure
there is no administration set or foreign object in the tubing channel.
2.
Turn the release tab (Figure 11-16A) clockwise until it stops and
push the tab into its socket. For triple channel pumps: Repeat the
steps above as needed for additional channels.
3.
Press the
Done
soft key to clear the alarm.
Note:
If the MTR is used following a channel failure to remove the
administration set, the pump cannot be powered off until the MTR
has been reset. A Reset Manual Tube Release pop-up message
will be displayed.
Note:
If three unsuccessful attempts to reset the MTR are made, a
channel failure occurs. The pump cannot be used until the MTR
is reset and the pump is powered off and back on.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...