9
Pump Specifications
9-2
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Range of Programmable Flow
Rates
Primary Infusion
• 0.1 to 99.9 mL/hr in 0.1 mL/hr increments
• 1 to 1200 mL/hr in 1 mL/hr increments
Secondary Infusion
• 0.1 to 99.9 mL/hr in 0.1 mL/hr increments
• 1 to 500 mL/hr in 1 mL/hr increments
Note:
Rate limits can be configured for values less than those stated above.
! WARNING !
There may be periods of no flow for flow rates less than or equal to 1mL/hr.
Volume to be Infused
• 0.1 to 99.9 mL in 0.1 mL increments (micro).
• 1 to 9999 mL in 1 mL increments (macro).
Note:
Volume to be infused limits can be configured for values less than those
stated above.
Patient Weight Range
(for dose modes based on
patient weight)
Programmable from 0.2 to 99.9 kg in 0.1 kg increments.
Programmable from 100 to 600 kg in 1 kg increments.
(Programmable from 0.44 to 99.99 lb in 0.01 lb increments.
Programmable from 100 lb to 1322 lb in 1 lb increments).
Note:
For each PERSONALITY feature set, patient weight range (in kg) and the
units for patient weight entry (lb or kg) can also be configured within the limits
specified above.
KVO
0.1 to 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate,
whichever is less.
Priming Rate
500 mL/hr.
Note:
To be used only when the administration set is not connected to patient.
Advance Air Rate to Advance
an Air Bubble
At the same rate programmed for the current primary or secondary infusion.
Air Bubble Setting
The air bubble setting is a configurable option.
The air sensor measures the accumulated amount of air detected over an amount of
solution delivered. The amount of delivered solution depends on the programmed
bubble size. The air alarm is triggered for a single air bubble greater than the set
threshold or an accumulation of air greater than the threshold. The alarm threshold
and accumulation volumes are given in the table below.
Air Bubble Setting
Accumulation Volume
25 microliters
0.83 mL
50 microliters
1.67 mL
100 microliters
3.33 mL
150 microliters
5.00 mL
Component
Description
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...