Ipump Pain Management System Operator’s Manual
5-27
CD07-19-E4-766
5.
Press
START
. The automatic upstream occlusion test is performed (see
“Upstream Occlusion Testing,” 5-18). During the test, the
Alert
LED flashes
red.
Upon successful completion of the test, the bolus starts, and the
Infusing
LED
flashes green.
NOTE:
The pump will not begin infusing until it has successfully passed the upstream
occlusion test. If an
UPSTREAM OCCLUSION
or
SET NOT PRIMED
alarm occurs, the
test must be repeated until it passes successfully. See Table 6-2, "Alarm Messages and
Responses," on pages 6-18 and 6-15 for more information.
TESTING
UP OCCLUSION
Example:
BOLUS INFUSING
1000.0 mg
6.
Wait until the bolus delivery is completed, or press
STOP
twice in 1 second to
stop the infusion.
Example:
BOLUS DONE
1000.0 mg
Action
Message Displayed