Ipump Pain Management System Operator’s Manual
4-8
CD07-19-E4-766
Air Detection
This setting controls the sensitivity
of the Air Detection feature.
OFF
LOW = 0.5 mL within 0.8 mL of fluid
HIGH = 0.1 mL within 0.16 mL of fluid
NOTE:
During the upstream occlusion
test, the Air Detection sensor is turned
on regardless of its configured setting.
See “Upstream Occlusion Testing” on
page 5-18 for more information.
OFF
The air sensor will measure the accumulated
amount of air detected over an amount of fluid
delivered. The amounts of delivered fluid
depend on the programmed air bubble size. The
air alarm is triggered for a single air bubble
greater than the set threshold or an
accumulation of air greater than the threshold.
! WARNING !
The air sensor detects and measures
the accumulated amount of air over an
amount of fluid delivered. However, the
pump may not detect all instances of
micro or “champagne” air bubbles.
Table 4-3
Configurable Options—Controls — continued
Controls
Available Settings
Factory Settings