Ipump Pain Management System Operator’s Manual
4-7
CD07-19-E4-766
PCA Button Status
REQUIRED
OPTIONAL
REQUIRED
NOTE:
In PCA and Basal+PCA modes,
the REQUIRED setting requires use of a
PCA button to request PCA doses. The
OPTIONAL setting allows the use of the
START
key as an alternative for
requesting PCA doses.
Preventive Maintenance Alert
0 (no alert) to 12 MONTHS
0 (no alert)
Upstream Occlusion Detection
ON
OFF
ON
NOTE:
When the pump is configured for
upstream occlusion detection, it
continuously detects for upstream
occlusions throughout an infusion. This is
in addition to the upstream occlusion
testing performed at the start of an infusion
(see “Upstream Occlusion Testing,” 5-18).
Table 4-3
Configurable Options—Controls — continued
Controls
Available Settings
Factory Settings
Oct02_0719E4766.book Page 7 Friday, August 3, 2007 1:46 PM