sound after the pump is started. If the cover cannot be closed and locked, check the tubing door to be certain
it is closed completely.
Changing the Tubing Set and/or Fluid Bag
1.
Follow the instructions given in “Loading the Tubing Set,” 3-14, and “Filling and Loading the Fluid Bag,”
3-16, to prepare and install the bag and tubing set.
2.
Complete the tasks in the “Action” column of the procedure “Changing the Prescription During Infusion,”
5-35, except select
FLUID VOLUME
at the
SELECT ACTION
prompt.
3.
Enter the correct fluid volume and prime the tubing set, if required (see “Prime the Tubing Set,” 5-16).
4.
When the pump displays the
PUMP READY
prompt, press
START
to restart the infusion.
! WARNING !
If the tubing is pinched by the closed tubing door, resistance to flow may increase and fluid
delivery to the patient may be compromised. If this occurs, open the bag cover, reseat the tubing
and check the tubing door for proper closure, close the bag cover, then restart the pump.
CAUTION
The time to detect occlusions increases proportionally with a decrease in flow rates. At flow
rates below 0.5 mL/hr, an occlusion may not be detected. Baxter recommends that extra care
be taken to ensure that neither the tubing nor the bag are pinched, twisted, or occluded.
! WARNING !
The tubing set MUST NOT be connected to the patient while priming.