Ipump Pain Management System Operator’s Manual
3-17
CD07-19-E4-766
Q
Inject the solution into the bag. If necessary, refill the syringe and repeat the process. (The bag will hold
approximately 100 mL or 250 mL, depending on the bag selected.)
Q
Remove all air from the bag by aspirating with the syringe, and then remove the syringe.
Q
If the bag is being used immediately, do not use the nonvented cap. Connect the luer-lock fittings
between the bag and the pump tubing set. (The bag must be connected to the shorter segment of the
tubing set.)
If the bag is being stored for later use, connect the nonvented cap to the bag after the bag is filled.
For Baxter Viaflex, IntraVia, or similar spike fluid bags:
Q
Add any additional drug/diluent to the bag using a syringe and needle appropriate for the injection port
of the bag. Mix or shake the bag to dilute the drug and solution appropriately.
Q
Remove all air from the bag by aspirating with the syringe, and then remove the syringe.
Q
Insert the tubing set spike into the outlet port of the bag. Ensure that the spike is inserted up to the ridge
on the spike.
3.
Remove the air from the remainder of the tubing set by following the procedure in “Prime the Tubing Set,”
5-16.
! WARNING !
To help prevent injuries, always follow your facility's policies and procedures when using and
disposing of needles.
Oct02_0719E4766.book Page 17 Friday, August 3, 2007 1:46 PM