Ipump Pain Management System Operator’s Manual
1-13
CD07-19-E4-766
Cautions
CAUTION
In the US, use of this pump is restricted to sale or use by, on the order of, or under the supervi-
sion of a qualified physician.
CAUTION
There are no internal user serviceable parts or adjustments.
CAUTION
When using the optional AC adapter, use earth-grounded AC outlets only. When grounding relia-
bility is in doubt, the equipment should be powered by its battery.
CAUTION
Variations in epidural catheter sizes can cause downstream occlusion alarms. If an occlusion
alarm occurs with no visible occlusion, change to a larger diameter and/or shorter catheter. If
occlusion alarms continue, contact your nearest authorized service center.
CAUTION
Do not use sharp objects to press keys.
CAUTION
The time to detect occlusions increases proportionally with a decrease in flow rates. At flow rates
below 0.5 mL/hr, an occlusion may not be detected. Baxter recommends that extra care be taken
to ensure that neither the tubing nor the bag are pinched, twisted, or occluded.
CAUTION
The pump can be configured with the upstream occlusion detection disabled. If this feature is dis-
abled, Baxter recommends that extra care be taken to ensure that neither the tubing nor the bag
are pinched, twisted, or occluded.
CAUTION
This pump has configurable options. Operating modes and input parameter selections may vary
as a function of the selected configuration.