Ipump Pain Management System Operator’s Manual
6-8
CD07-19-E4-766
RELEASE THE
<key name>
KEY
Situation: A key on the key pad has been pressed continuously for 3 minutes, or the
key is stuck.
Action:
1.
Release the stuck key.
2.
If this alert occurs and the key is not being pressed intentionally, there may be a
mechanical or electronic fault in the key. Do not use the pump. Take the pump
out of service and contact your nearest authorized service center.
RELEASE PCA
BUTTON
Situation: The PCA button has been pressed continuously for 3 minutes.
Action:
1.
Release the PCA button.
2.
Advise the patient to press the PCA button briefly when requesting a PCA dose.
3.
If this alert occurs and the PCA button is not being pressed intentionally, there
may be a mechanical or electronic fault in the PCA button. Do not use the
pump. Take the pump out of service and contact your nearest authorized service
center.
Table 6-1
Alert Messages and Responses — continued
Alert Message
Situation/Action