Ipump Pain Management System Operator’s Manual
5-46
CD07-19-E4-766
10.
The HOURLY HISTORY group of screens is displayed only if PCA doses were
allowed at some time during the therapy. For each 24-hour period, three screens
per hour are generated. At the
HOURLY HISTORY
message, press the
⇒
key to
scroll through the following screens (or press
↑
to proceed to the EVENT
HISTORY group messages):
HOURLY HISTORY
10.1
The hour of the dosage.
Example:
11:30 - 12:00AM
10.2
The number of PCA doses administered and the number requested
during the hourly period.
Example:
0004 INJECTIONS
0004 ATTEMPTS
10.3
The cumulative total infused at the end of the hourly period, including
any bolus, PCA, and Basal+PCA infusions.
Example:
0031.5 mL GIVEN
AS OF 12:00AM
11.
The EVENT HISTORY group displays a chronological list of events that
occurred during the therapy. This group begins with the message
EVENT
HISTORY
and ends with the message
END OF HISTORY
. (Press
↑
to proceed
directly to
END OF HISTORY
.)
EVENT HISTORY
Action
Message Displayed
Oct02_0719E4766.book Page 46 Friday, August 3, 2007 1:46 PM