7 Precautions/ Safety Instructions
OccluSense
®
is a medical device meant for professional use only (see chapter 3).
The OccluSense
®
-System must be used with the supplied or prescribed accessories only.
It is not allowed to connect any part of the OccluSense
®
-System to an external DC power
source. The power supply of the system has to be carried out exclusively with the
supplied charging accessory.
Before using the OccluSense
®
-System, components have to be checked for external
damages (visual inspection). Damaged components must not be used.
The OccluSense
®
-System must not be altered without the manufacturer's permission.
The sensor must not be modified (e.g. cut to size).
The handheld is designed for battery operation only. Mains operation is not intended
and may result in serious injury of the patient or the practitioner.
The use of non-rechargeable batteries is not permitted and may result in serious injury
of the practitioner or destruction of the handheld when attempting to recharge it.
The handheld must be disconnected from the charging station during use.
The charging station has to be located outside the patient´s vicinity.
The charging station must be positioned in such a way that the plug-in power supply
unit is in easy reach and the separation from the mains can be accomplished without
difficulty.
The charging station can be disconnected from the mains by pulling the plug-in power
supply out of the socket.
All parts of the OccluSense
®
-System are not thermo-disinfectable or autoclavable and
are not delivered sterile.
Only cleaning agents and disinfectants as described in chapter 21 „Cleaning and
disinfection“ are allowed.
The charger is maintenance-free.
The treatment light should not be directed towards the handheld during treatment to
ensure good readability of the display.
The OccluSense
®
-System is not intended to serve as the sole basis for diagnosis / the-
rapy decisions.