7.3. Switching ON and OFF

Caution:

 Connect the magnetometer before switching on the SCU1, as connecting a live 

cable to the magnetometer may cause damage. Similarly, switch off the SCU1 before 
disconnecting the magnetometer.

Caution:

 The SCU1 should be switched on and off using the IEC socket switch (Figure 1, 

item 4) (“0” = OFF, “1” = ON).

Note: 

For best results, after switching ON, leave the SCU1 for 10 minutes for the internal 

temperature to stabilise, before performing any measurements.

7.4. Using the SCU1 Controls

7.4.1. Offset Controls

The offset controls apply a DC offset value to the magnetometer output voltage, to generate a DC 
adjusted signal voltage. This resulting voltage will be amplified by the gain controls, to form the 
final SCU1 conditioned output signal.

The offset voltage range limits are -10 to +10V.

The offset control includes a coarse adjust, to set an approximate level, and a fine adjust to 
achieve the exact offset level desired. A polarity switch selects whether the offset is added to or 
subtracted from the magnetometer output, or set to OFF (zero offset).

Offset is generally used where an AC magnetic field being investigated is superimposed on a 
background static (DC) field. The DC signal can be eliminated from the SCU1 conditioned output 
by applying offset. The conditioned output will then contain only the AC signal, which can be 
investigated at high resolution by using gain (amplification).

Example:

 to remove the background DC field from the X channel:

1.  Set the X channel gain to 1, and the HPF to DC.

2.  Set the X channel offset polarity switch to the centre position (no offset).

3.  If the X channel meter shows a + reading, set the offset polarity to -; if meter shows -, set 

offset polarity to +.

4.  Adjust the offset controls until the meter reading decreases to zero.

7.4.2. Gain Controls

The gain controls apply amplification to the DC adjusted signals from the offset stage, to 
generate the SCU1 conditioned analogue outputs.

Содержание SCU1

Страница 1: ...Operation Manual for SCU1 Signal Conditioning Unit...

Страница 2: ...1 Back Panel Connections and Controls 7 5 2 Front Panel Displays and Controls 8 6 Installing the SCU1 10 6 1 Location of the Equipment 10 6 1 1 Environmental Precautions 10 6 1 2 Mounting 10 6 1 3 Ori...

Страница 3: ...Controls 13 7 4 2 Gain Controls 13 7 4 3 Low Pass Filter LPF 14 7 4 4 High Pass Filter HPF 14 7 5 Using the SCU1 Outputs and Displays 15 7 5 1 Unconditioned and Conditioned Outputs 15 7 5 2 Display Pa...

Страница 4: ...aution is ignored Identifies items that must be disposed of safely to prevent unnecessary damage to the environment Note Provides useful supporting information on how to make better use of your purcha...

Страница 5: ...via the following outputs unconditioned analogue outputs signals direct from the magnetometer conditioned analogue signals the magnetometer signals modified by the SCU1 analogue conditioning digital d...

Страница 6: ...t 3 Gain Finally the DC adjusted output signal can be amplified by a variable gain stage to produce the conditioned output signal Offset and gain are controlled independently for the X Y and Z channel...

Страница 7: ...ter Output Type Selector Switch The SCU1 can be used with magnetometers that have unbalanced single ended outputs or balanced differential output signals This switch sets the SCU1 to operate in unbala...

Страница 8: ...HPF Control The SCU1 can apply high pass filtering to the magnetometer output signals This control sets the 3dB point frequency for the HPF 9 Magnetometer Supply Voltage Control This control sets the...

Страница 9: ...et voltage is subtracted Table 1 Offset switch functions 12 Offset Coarse Controls one for each channel 13 Offset Fine Controls one for each channel These rotary controls set the offset DC voltages ap...

Страница 10: ...ckets 6 1 3 Orientation The SCU1 can be orientated horizontally or vertically 6 1 4 Temperature To minimise temperature induced drift effects position the SCU1 in a constant ambient temperature out of...

Страница 11: ...ometer Supply Voltage Before switching the equipment ON set the appropriate voltage for your magnetometer using the front panel control Figure 2 item 9 For all Mag648 and Mag 03RC magnetometers use th...

Страница 12: ...ometer with a range of 60 T and an analogue output of 3V has a scale factor of 20 To convert voltage to field strength 1 Divide the output voltage by the gain control setting then 2 Multiply by the sc...

Страница 13: ...e adjust to set an approximate level and a fine adjust to achieve the exact offset level desired A polarity switch selects whether the offset is added to or subtracted from the magnetometer output or...

Страница 14: ...ing AC fields or very slow changes to a background DC field In these situations any DC signal present would normally be removed by using offset This allows the level of gain to be set as high as possi...

Страница 15: ...C content of the signals need to be measured e g with an oscilloscope meter or an A to D system or when a voltage output is required for connection for example to a feedback or control system Note The...

Страница 16: ...artington com or telephone the Bartington Instruments service team on 44 0 1993 706565 Fault Possible cause Solution Panel meters are off Power fault Check SCU1 is connected to a mains supply Check th...

Страница 17: ...truments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 9 3 Cleaning WARNING Disconnect the electrical supply before performing any cleaning oper...

Страница 18: ...NSTRUMENTS Page 18 of 19 OM2441 1 10 End of Life Disposal This product electrical and electronic equipment should not be placed in municipal waste Check local regulations for disposal of electronic pr...

Страница 19: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Отзывы: