background image

1

 

Rev. 3.1 (08.2020)   

 

        

 

 

Operating Manual for 
Micropump mp6-series 

 

 

Содержание mp6

Страница 1: ...1 Rev 3 1 08 2020 Operating Manual for Micropump mp6 series...

Страница 2: ...specifications 5 Technical Data of the mp6 hyb1 5 Technical Data of the mp6 liq1 2 6 Technical specifications mp6 pi1 7 Technical specifications mp6 pp1 9 Technical specifications mp6 gas1 10 Final in...

Страница 3: ...Low Voltage Directive 2014 35 EU Description of functions The micropumps have been developed for the transport of gases or liquids The controllers have been developed for operating the micropumps Bar...

Страница 4: ...ditions including vapors is not permitted Staff selection and qualification All work in connection with the installation assembly commissioning decommissioning disassembly operation servicing cleaning...

Страница 5: ...in typ flow rate p 0 8 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 Gases air typ volume flow p 0 20 ml min 5 typ back pressure p 0 80 mbar 1 16 psi 5 1 Typical values Values can vary under appl...

Страница 6: ...Please contact us for more information Please find more information concerning the controller and the equipment in the corresponding manuals mp6 liq Order code mp6 liq Pump type piezoelectric diaphrag...

Страница 7: ...1 1811 x 0 5906 x 0 1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 50 mW 5 Self priming...

Страница 8: ...Tygon tubing 1 3 mm inner diameter 3 Conditions Suction pressure 10 mbar DI water settings mp x 100 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions...

Страница 9: ...ter Controllable flow range8 8 l min 4000 l min typ flow rate p 0 4 ml min 4 typ back pressure p 0 500 mbar 7 25 psi 4 1 Typical values Values can vary under application conditions Content is subject...

Страница 10: ...1498 in Weight 2 g Fluidic connectors barbed tube clip outer diameter 1 9 mm length 3 5 mm 2 Electric connector flex connector 1 25 mm pitch Power consumption 150 mW 5 Self priming yes 3 Pumping medi...

Страница 11: ...er 3 Conditions air room temperature 23 C settings mp x 300 Hz 250 V SRS the max volume flow will be reached after a few minutes of operation time 4 Conditions gases room temperature 23 C mp x 300 Hz...

Страница 12: ...rating the micropump In this chapter the operation of the micropump during the evaluation is described This chapter provides information on the proper connection of the pump with tubing and electrical...

Страница 13: ...mp6 series to the cable The pumps can be connected via a FCC connector The layout of the connector and pin assignment of the pumps is as shown below Each piezo P1 P2 has a single lead for the negative...

Страница 14: ...n top or bottom of the pump the Molex connector is not water tight additional sealing with e g silicone will be necessary fixation of the Molex connector in the final assembly is recommended The Molex...

Страница 15: ...neral for driving the pump with air the following points should be taken into account Higher frequencies should be used compared to pumping water Large amplitudes respectively driving voltages should...

Страница 16: ...uencies The performance depends on the environmental conditions Combination of micropumps The micropumps can be combined to achieve higher flow rates or pressure than the single unit can generate With...

Страница 17: ...l 49 231 47730 500 Fax 49 231 47730 501 Visit our Website www bartels mikrotechnik de for further information on applications Tutorials and helpful answers to frequently asked questions can be found i...