P-600 SIC Analyzer Manual
Chapter IV: Programming
29
Alarm Critical
–
This type of alarm indicates Analysis has stopped because one of the
critical conditions has been detected.
Maintenance
–
Activated when the number of measurement cycles has exceeded the
Max Cycles setting.
Stream 1
–
stream 1 is being measured.
Stream 2
–
stream 2 is being measured.
Come Read
–
Activated for a programmed period of time upon the completion of a
measurement cycle.
In Validation
–
validation cycle is being performed.
Valid. Pass
–
validation cycle passed.
•
Conditions
—
This designates whether the alarm relay will energize or de-energize when an
alarm condition is detected. When Normal is selected, power will be applied to the relay when an
alarm condition is detected; when Failsafe is selected, power will be removed from the relay when
an alarm condition is detected.
Communications Setup
The Communications submenu allows you to configure the Analyzer’s Serial and Modbus settings.
Serial Port C
This is the Analyzer’s RS-232 serial communication port.
•
Mode
—
This menu item determines what data will be output through the RS-232 port. The
choices are Result, Data, and None.
•
Rate
—
This is the baud rate used for serial communications through this port (9600 / 19200 /
38400).
Modbus