P-600 SIC Analyzer Manual
Chapter IV: Programming
26
NOTE: If Salt in Crude graphing is enabled, the graph is updated at the
end of each measurement.
Alarm Settings
This submenu screen is used to program the Analyzer’s alarms.
Use the Index key to move to the desired menu item. Use the Up/Down Arrow keys to change the displayed
setting or value. The unit of measure used for these values is set in the Other Settings submenu.
Warning Alarms
Those types of alarms do not stop the analysis.
“Result out of range” warning alarm is activated when the
measured concentration exceeds the indicated value.
•
Low
—
This menu item allows you to establish a low alarm level. Salt concentration
measurements below this value generate a warning alarm signal/message. Analysis will continue.
•
High
—
This menu item allows you to establish a high alarm level. Salt concentration
measurements above this value generate a warning alarm signal/message. Analysis will
continue.
Critical Alarms
Those types of alarms stop the analysis. Programmable critical alarm is a maximum cell temperature.
•
Maximum Temperature
—
This menu item allows you to establish a maximum allowable cell
temperature. Cell temperatures above this value generate a critical (fatal) alarm signal/message
and stops analysis.
•
Rem Stdby
—
Enables or Disables Alarm Critical when in Remote Standby mode.