Maintenance
79
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
● Push the measuring cell downwards and out of the middle measuring
chamber section. Attention: never press the “tulip” but press carefully
be
side the “tulip” to the measuring cell by using a screw driver or a similar
tool.
6.3.5
Changing the humid temperature
sensor
● Detach the knurled nut from the sensor mount and carefully extract the
sensor.
● Insert the new sensor and tighten the knurled nut.
● After reassembling the measuring chamber, connect the sensor’s
connector in the enclosure and tighten the knurled nut.
● Sensor calibration see section 5.2.4.7.