6

G

ENERAL

S

PECIFICATIONS

Heating Specifications

Top Plate Surface - Solid Aluminum

Temperature range: 1°C - 300°C (34°F - 572°F)*

* These plates do not cool. The minimum temperature is 1°C if used in a cold room below 1°C.

Heat-up time to within 5°C of maximum temperature (unloaded top plate).

5 minutes

Accuracy of the temperature display vs. the actual average temperature 

± 10.0°C

of a 2” diameter of setting area at the center of the top plate (setpoint 
100°C unloaded). 

- Temperature stability at the center of the top plate surface 

± 1.0°C typical

(@ 100°C unloaded).

Accuracy of remote probe at user selected calibration temperature after

±0.5 typical

calibration procedure.

- Temperature stability using remote probe  (500 ml of water 

±0.5°C

in a 1000 ml flask at 70°C)

Stirring Speed Specifications

Speed Range 50 to 1200 RPM 
(Maximum speed is dependent on the viscosity of the solution)

Stability of the stirring speed setpoint (600 ml of water in a 1000 ml glass flask above 200 rpm)

± 2.0%

Accuracy of the stirring setpoint (600 ml of water in a 100 ml glass flask above 200 rpm)

± 2.0%

Max Recommended Flask Size

Max Weight on Top Plate

4 liters

25 lbs

Содержание SUPER-NUOVA HP133420-33

Страница 1: ...S LIST Model Voltage Description HP133425 120 7X7 Hot Plate HP133420 33 220 240 7X7 Hot Plate HP133424 100 7X7 Hot Plate SP133525 120 7X7 Stirring Hot Plate SP133520 33 220 240 7X7 Stirring Hot Plate...

Страница 2: ...Metal Vessels and Sand Baths 16 Setting the Timer 16 Power Interruption Protection 17 Using the Probe with the Temperature Display for Stirring only Applications 17 RS232 Output 18 Using HyperTermina...

Страница 3: ...ating volatile materials top surface and element can reach the Flash Point Temperature of many chemicals These plates are not explosion proof Fire or explosion may result Unit contains components whic...

Страница 4: ...ols to below 50 C 9 Do not leave an active probe out of the fluid This may cause uncontrolled heating of the fluid on the hot plate and unintentional boiling or an explosion could occur 10 Localized h...

Страница 5: ...C p m e T x a M C 0 0 3 F 2 7 5 C 0 0 3 F 2 7 5 C 0 0 3 F 2 7 5 Specifications s n o i s n e m i D l l a r e v O 3 3 0 2 5 3 3 1 P S 4 2 5 3 3 1 P S 5 2 5 3 3 1 P S m c n i h t d i W 8 0 2 2 8 8 0 2 2...

Страница 6: ...rature stability at the center of the top plate surface 1 0 C typical 100 C unloaded Accuracy of remote probe at user selected calibration temperature after 0 5 typical calibration procedure Temperatu...

Страница 7: ...nt for measurement control and laboratory use Part I General Requirements Safety EN 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use Part I General Requi...

Страница 8: ...o display the actual temperature of heat sensitive solutions during stirring only functions The top plate on the SUPER NUOVA units are solid aluminum and are suitable for use with glass or metal vesse...

Страница 9: ...dset into a proper ly grounded electrical outlet of correct voltage and current handling capacity Unpacking and Installation Warning Use a properly grounded electrical outlet of correct voltage and cu...

Страница 10: ...ON UTION HOT TOP HOT TOP 1 2 4 13 7 8 9 6 3 12 14 10 16 15 11 5 Use only a No 2 pencil to gently mark up writable preset surface SUPER NUOVA Stirring Hot Plate Control Panel PRESETS PRESETS POWER POWE...

Страница 11: ...n 4 Power Button 5 Writable Preset Surface 6 Preset Buttons 7 External Probe Indicator Light 8 Set Indicator Light 9 Knob 10 Timer 11 Timer Indicator Lights 12 Speed Set Button 13 Speed Display 14 Pau...

Страница 12: ...e in safety due to the redundan cy of the circuits It is possible to heat certain types of metal vessels and sandbaths on the SUPER NUOVA without damaging the aluminum top The electronic stirring spee...

Страница 13: ...lit when in SET mode Once the unit begins stirring the display will register ACTUAL speed until the setpoint has been reached To check the setpoint press SET under the STIR display and your setpoint...

Страница 14: ...SUPER NUOVA plate will accept any ungrounded Type K thermocouple probe however the accuracy of probes other than those tested and specified may not reach the probe accuracy rating designated To set th...

Страница 15: ...temperature When using a probe with the SUPER NUOVA plate it is rec ommended that a clamp on a support rod be used to hold the probe in the solution To ensure accurate probe readings as much of the pr...

Страница 16: ...ce Heating Metal Vessels and Sand Baths Metal vessels and sand baths cannot be heated on most solid aluminum tops because of the tendency for the metal and the sand to reflect heat back into the top e...

Страница 17: ...once the facility power is restored Using the Probe with the Temperature Display for Stirring only Applications On stirring hot plate models the probe can also be uti lized to monitor the actual temp...

Страница 18: ...are given in degrees Celsius When the unit is first powered on displays turn on a header will print indicating the columns In order to communicate with the SUPER NUOVA the com puter connected to it ne...

Страница 19: ...to hot plate communi cations cable Connect it from the back of the hot plate to the communications port desired To Setup HyperTerminal 1 From the START menu of Windows select ACCESSORIES COMMUNICATIO...

Страница 20: ...Select TRANSFER CAPTURE TEXT Select the Folder and File name you would like to use to refer back to this data example C SuperNuova Test1 txt It is important to use the txt extension so that when you d...

Страница 21: ...b o t o t t i u c r i c P T O e h t w o l l a n a c d e t i m i l n U f i r e w o p y a l e r t c e n n o c s i d y l l a c i t a m o t u a t l u a f e d e h T d e d e e c x e s i g n i t t e s P S O...

Страница 22: ...below 1 Plug the unit into the appropriate power but do not turn it on displays should be blank unless Hot Top Warning System is indicating a hot sur face enter the Service Menu by pressing and holdi...

Страница 23: ...ependent temperature source such as a bath is not available The hot plate itself is used to supply the bath Although an independent and accurate method of determining temperature is still required Thi...

Страница 24: ...d and you may remove your finger from the POWER key The display will change to OSP 3 Rotate the center knob dial one position clock wise and the display will read CAL This is the calibration menu Pres...

Страница 25: ...sound and the SET indicator under the HEAT display will begin flashing to let the user know that the display is ready to be adjusted using the center knob Although it may be desirable to wait longer t...

Страница 26: ...t be connected before continuing The SET indicator under the HEAT display will be flashing to let the user know that the display is ready to be adjusted using the center knob 5 Adjust the HEAT display...

Страница 27: ...ress the POWER key and the unit will turn OFF When the unit is turned back ON it will be back to nor mal operation Timer Shutdown End This feature is available for all units but not all sub menu optio...

Страница 28: ...that is not a concern or a different material of probe is chosen then unlimited is a safe choice To modify view the Probe Temperature Limit follow the instructions given below 1 Plug the unit into the...

Страница 29: ...r from the POWER key The display will change to OSP 2 Rotate the center knob until the display reads Pr Press the SET key under the HEAT display to accept and the display will change to the cur rent t...

Страница 30: ...y under the HEAT display again 3 The unit will return to the previous menu Select another feature to change or press the POWER key again to return to the off mode Error Disable Err The purpose of this...

Страница 31: ...off Factory Defaults deF This feature is available for all units and is designed to restore the control to the default factory settings NOTE The Model Selection feature will need to be re visited aft...

Страница 32: ...ding the POWER key After about 3 seconds a single beep will be heard and you may remove your finger from the POWER key The display will change to OSP 2 Rotate the center knob until the display reads S...

Страница 33: ...s NOTE Special software and an RS232 cable are required to complete this proce dure Complete instruction will be included with the firmware upgrade General Cleaning Instructions Unplug unit and remove...

Страница 34: ...c e n n o c e b o r p l a n r e t x E s d r a w k c a b o t n i e b o r p l a n r e t x e e c a l P n o i t u l o s g n i t t e s P T O e s a e r c n I e b o r p l a n r e t x e e s a e r c n I e s n...

Страница 35: ...o r o t o m e c a l p e R r e d o c n e 1 2 E y r o m e m h s a l f a t a d d e t p u r r o C a t a d g n i r u d e r u l i a f m u s k c e h C y r e v o c e r h s a l f r o e c i v r e S r e m o t s...

Страница 36: ...36 Exploded Views 6 5 7 8 9 10 11 22 13 14 12 16 15 2 1 3 4 7x7 Stirring Hot Plate Exploded View...

Страница 37: ...37 EXPLODED VIEWS 21 18 19 20 17 Note routing of leads and thermocouple Number located this side 7x7 Stirring Hot Plate Top Assembly...

Страница 38: ...38 EXPLODED VIEWS 7 x 7 Hot Plate Exploded View 5 6 7 8 10 22 12 13 16 1 2 3 4...

Страница 39: ...39 EXPLODED VIEWS 21 17 18 19 20 7 x 7 Hot Plate Top Assembly...

Страница 40: ...CEX373 ELX51 INLET ELEMENT CONTROL PCB DESCRIPTION 7x7 EL1 120V ELEMENT ELX50 EL1 220 240V ELEMENT ELX52 PC1 100 120V CONTROL PCB C2 EXT PROBE CEX376 C3 RS232 WH1318X1 F1 220 240 ONLY FUSE 261575 S2 S...

Страница 41: ...D SENSOR COMPONENT CHART CEX373 ELX51 MT1318X2 INLET ELEMENT MOTOR CONTROL PCB DESCRIPTION 7x7 EL1 120V ELEMENT ELX50 EL1 220 240V ELEMENT ELX52 MT1 100 120V MOTOR MT1318X1 PC1 100 120V CONTROL PCB C2...

Страница 42: ...211 Membrane Switch 8 KBX105 KBX105 Knob 9 MGX24 Magnet 10 CEX373 CEX373 Power Entry Module 11 SC1313X1 Speed Sensor 12 CRX106 CRX106 Cord Set 100 V 120V 12 CRX104 CRX104 Cord Set 220 240V 13 KBX106 K...

Страница 43: ...ion 1007 3BLS Non Mercury Thermometer 1 to 201 C Range 76 mm immersion 615 3SC Mercury Teflon Coated Thermometer 10 to 200 C Range 76 mm immersion 260CW 3BLS Non Mercury Thermometer 10 C to 260 C Rang...

Страница 44: ...his unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request then contact ou...

Страница 45: ...45...

Страница 46: ...46...

Страница 47: ...47...

Страница 48: ...ing use inconsistent with written operating instructions for the product mishan dling contamination overheating modification or alteration of the product by any customer or third party or iii use of r...

Отзывы: