22
General (con't)
Problem
Causes
Solution
Steam at Condenser Vent.
A slight wisp of steam exiting
at the condenser vent is
normal. Gases in the steam are
also being removed.
If there is an excessive amount
of steam leaving the
condenser vent, increase the
cooling water flow rate.
No Steam at Condenser Vent.
If no steam is visible, decrease
cooling water flow rate.
Water Pressure Variations.
Pressure fluctuations in your
tap water line will cause erratic
flow in cooling water and may
cause water level in boiler to
drop.
Have a plumber install a
pressure regulator in your tap
water line and regulate to 20
-
25 psi.
Output Less Than 1 Liter/Hr.
Heaters will produce in excess
of 1 L/hr. of distilled water
when run at the rated voltage.
At lower voltages, output will
drop accordingly.
Also see "Steam at
Condenser Vent" as excess
steam at this point will reduce
output.
M
AINTENANCE AND
S
ERVICING