Replacing Fuses

There are two fuses located at the rear of the
unit. The main fuse for the unit is a 20 Amp,
fast-acting, type KLK midget fuse. A 0.5 Amp,
5x20mm, fast-acting fuse protects the unit’s low
power electronics. If this fuse blows, use a slot-
ted screwdriver to remove the old fuse and
install a new 0.5 Amp fast-acting fuse. 

General Cleaning Instructions

Wipe exterior surfaces with a lightly dampened
cloth containing a mild soap solution.

8

Maintenance and Servicing

Содержание SET-TEMP limit

Страница 1: ...SET TEMP limit OPERATING INSTRUCTIONS 1 BARNSTEAD THERMOLYNE CORPORATION LT1153X2 6 24 99...

Страница 2: ...of Operation 5 Installation 6 Temperature Probe Connection and Placement 6 Connecting your Heater and Other Temperature Controls 6 Operation 7 Maintenance and Servicing 8 Replacing Fuses 8 General CL...

Страница 3: ...ting tapes hot plates ovens baths and other general heat ing equipment General Usage Do not use this product for anything other than its intended usage Warnings To avoid personal injury 1 Refer servic...

Страница 4: ...heaters up to 1800 watts 100 120 VAC Shipping weight 4 4 lb 2 kg Selected probe will dictate actual temperature range within range specified 4 General Specifications Environmental Conditions Operatin...

Страница 5: ...off power to laboratory heaters to guard against thermal runaway when the temperature sensed by a thermo couple exceeds the dial setting Heating will not resume until the RESET key is pressed 5 Princ...

Страница 6: ...ood idea to have the tip of the probe resting on the inside bottom of the flask Then should the liquid boil away the sensing probe will still be effectively heated by the man tle and SET TEMP LIMIT wi...

Страница 7: ...ure scale is in degrees C and degrees F 3 Press the reset key to energize SET TEMP LIMIT s receptacle Note that each time SET TEMP LIMIT looses power or is turned off the RESET key must be pressed to...

Страница 8: ...fuse A 0 5 Amp 5x20mm fast acting fuse protects the unit s low power electronics If this fuse blows use a slot ted screwdriver to remove the old fuse and install a new 0 5 Amp fast acting fuse General...

Страница 9: ...mm so it is compatible with common ther mometer adapters clamps and apparatus The glossy quartz sheath may be cleaned more thoroughly than metals Teflon or other glasses Usable up to 500 C Each therm...

Страница 10: ...ROBE 1 4 diameter x 12 long Type J Thermocouple 8005 GLASS PROBE 1 4 diameter x 18 long Type K Thermocouple 8007 GLASS PROBE 1 4 diameter x 24 long Type K Thermocouple 8403 Bendable s s probe 1 16 dia...

Страница 11: ...whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your re...

Страница 12: ...wn discre tion Heating elements however because of their susceptibility to overheating and contam ination must be returned to our factory and if upon inspection it is concluded that failure is not due...

Отзывы: