LT1465X3 • 3/14/06
Imperial III General Purpose
Incubators
Series 1464, 1465, 1466
Model No.
Gravity Convection:
302, 302-1305, 305-1 310, 310-1
Mechanical Convection:
305M, 305M-1310M, 310M-1
Страница 1: ...LT1465X3 3 14 06 Imperial III General Purpose Incubators Series 1464 1465 1466 Model No Gravity Convection 302 302 1 305 305 1 310 310 1 Mechanical Convection 305M 305M 1 310M 310M 1...
Страница 2: ...r 10 Auto Tune 11 Temperature Calibration 12 Troubleshooting 14 Maintenance 15 Checking Controls 15 Cleaning 15 Care and Cleaning of Stainless Steel 15 The Alloy Called Stainless 15 Stainless Guidelin...
Страница 3: ...ithout disturbing the interior environment Slide out shelving can be positioned to meet user requirements A PID microprocessor based controller maintains cham ber temperature An over temperature safet...
Страница 4: ...ly to lead to improper opera tion erroneous calibrations and possible damage to the equipment Do not attempt operation without this informa tion Safety Information Warning Warnings alert you to a poss...
Страница 5: ...60 2 5 600 1 electrical outlet 120V 500W total load 310M 1 230 240 50 60 2 5 600 1 electrical outlet 120V 500W total load Temperature Range Slightly above ambient to 65 C Shipping Weight 302 Models 7...
Страница 6: ...ing 15 C to 40 C 20 to 80 relative humidity non condensing Installation category II overvoltage in accordance with IEC 664 Pollution degree 2 in accordance with IEC 664 Altitude Limit 2 000 meters Sto...
Страница 7: ...tected from drafts strong air currents and large fluctuations in ambient tem perature It must be located near an electrical outlet that meets the unit nameplate requirements Allow clearance around the...
Страница 8: ...procedure for the opposing bracket making sure it is on the same level as the first bracket count the slots 4 Install other shelf brackets in the same way then slide shelves into place UNPACKING AND I...
Страница 9: ...ER TEMPERATURE THERMOSTAT This hydraulic thermostat is set by the operator to back up the control thermostat and safeguard the incubator Start Up Insert a flat blade screwdriver into the panel opening...
Страница 10: ...only at temperatures above the controller set point Temperature Controller 1 CONTROLLER SELF TEST When the incuba tor is powered up the controller will display 8888 along with the three decimal point...
Страница 11: ...parameters to achieve optimal temperature control It is not necessary to run the auto tune program when setting up the oven However if the temperature appears to be unstable the auto tune program can...
Страница 12: ...through the glass door 2 Press and hold the STAR key and using the UP or DOWN arrow key adjust the set point to the desired temperature 3 Allow the unit to run for at least 30 minutes 4 The controlle...
Страница 13: ...arrow keys together until the display again indicates the chamber temper ature If the procedure was done correctly the controller display should now agree with the thermometer reading to within 0 5 C...
Страница 14: ...A R E P O e r u t a r e p m e t r e v o e h t e t a t o r r e v i r d w e r c s a g n i s U t s e h g i h o t e s i w k c o l c y l e t e l p m o c m e t s t a t s o m r e h t p u t a e h t o n s e o...
Страница 15: ...faces Wipe exterior with soft damp cloth as necessary to remove dust fingerprints or other smudges Care and Cleaning of Stainless Steel WARNING Electrolysis can damage stainless steel This occurs when...
Страница 16: ...ing a bath or incubator ALWAYS add 2 to 40 ppm 20 to 40 mg liter disodium phosphate or sodium bicarbonate adjusting dosage to provide a pH value of 7 to 9 If not available use clean aerated soft tap w...
Страница 17: ...aqueous solution is kept within the range of 7 to 9 These are available through laboratory distributors Be sure to CONFIRM that they are not harmful to stainless steel Do not use any metallic pads Ins...
Страница 18: ...ust and scale is loosened Immediately follow with a clean water rinse Citric acid based cleaners Bathroom tub and tile cleaners A mixture of 20 nitric acid and 1 5 hydrofluo ric acid or hydrochloric a...
Страница 19: ...bulb out of the bracket and through the top hole to completely remove the thermostat 5 Install new thermostat by reversing above proce dure Switch or Status Lamp Replacement 1 Power switch and status...
Страница 20: ...sing a Use both hands to firmly grip each side of the controller bezel b Press on the bezel side grips until the bezel tabs release c Slowly pull controller from housing 4 To install new factory confi...
Страница 21: ...120 V Models 305 310 420 036 00 Receptacle 240 V Models 305 1 310 1 420 214 00 Shelves Models 302 302 1 2 810 353 01 Models 305 305 1 3 310 310 1 6 810 352 01 DESCRIPTION PART NUMBER Status Lamp Base...
Страница 22: ...R SWITCH 440 359 00 440 359 00 440 359 00 440 359 00 440 359 00 MTR FAN MOTOR MT1465X1 MT1465X1 MT1466X1 TPS TEMPERATURE SENSOR 410 632 00 410 632 00 410 632 00 CTR CONTROLLER CN71X134 CN71X134 CN71X1...
Страница 23: ...N MOTOR MT1465X2 MT1466X2 TPS TEMPERATURE SENSOR 410 632 00 410 632 00 410 632 00 410 632 00 410 632 00 CTR CONTROLLER CN71X134 CN71X134 CN71X134 CN71X134 CN71X134 CN71X134 HLP AMBER LAMP PL1365X1 PL1...
Страница 24: ...rom whom you pur chased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle you...
Страница 25: ...ead International for repair or credit that is contaminat ed with or has been exposed to potentially infectious agents or radioactive materials No product or component will be accepted without a Retur...
Страница 26: ...26...
Страница 27: ...27...
Страница 28: ...further force or effect if there is any i repair made to the product by a party other than BARNSTEAD or its duly authorized service repre sentative ii misuse including use inconsistent with written o...