General Purpose Aquabaths

Model No.

Digital Water Baths:

Basic Water Baths:

18002, 18002-1CE

18000, 18000-1

18007, 18007-1CE

18005, 18005-1 

18012, 18012-1CE

18010, 18010-1 

18022, 18022-1CE

18020, 18020-1  

18052, 18052-1CE

18050, 18050-1 

18102, 18102-1CE

18100, 18100-1 

18802, 18802-1CE

18800, 18800-1 

18902, 18902-1CE

18900, 18900-1

Teflon® Coated:

18202, 18202-1CE
18212, 18212-1CE

057-533-00 • 7/21/05

Содержание General Purpose Aquabath 18002

Страница 1: ...18000 18000 1 18007 18007 1CE 18005 18005 1 18012 18012 1CE 18010 18010 1 18022 18022 1CE 18020 18020 1 18052 18052 1CE 18050 18050 1 18102 18102 1CE 18100 18100 1 18802 18802 1CE 18800 18800 1 18902...

Страница 2: ...th Temperature is Maintained by Microprocessor Controller 11 Setting the Temperature Digital Water Baths 11 Temperature Calibration Digital Water Baths 11 Controls Basic Water Baths 12 Water Bath Temp...

Страница 3: ...ine is dry in the event of accidental spilling or splashing of liquids clean up and or neutralize the spilled liquids before continuing Do not use equipment for other than its intended purpose Use onl...

Страница 4: ...the chamber bottom helps distribute the heat uniformly and provides a level surface for a variety of glassware Its removal adds an extra inch of depth to the bath Heat is controlled from slightly abov...

Страница 5: ...e Range Slightly above ambient to 100 C with cover included as standard equipment Slightly above ambient to 65 C without cover Uniformity 0 2 C Control 0 15 C Specifications Basic Water Baths Electric...

Страница 6: ...2 3 1 5 7 8 W x 5 3 8 D x 5 H 8 3 8 W x 7 7 8 D x 9 H 18052 1CE 15 x 14 x 13 21 x 20 x 23 18102 22 0 11 5 8 W x 19 1 2 D x 5 H 13 3 8 W x 21 1 4 D x 9 H 18212 30 x 50 x 13 34 x 54 x 23 18802 6 7 5 7 8...

Страница 7: ...x 11 5 8 D x 2 1 2 H 7 1 2 W x 14 1 2 D x 7 H 18020 1 15 x 30 x 6 19 x 37 x 18 18050 3 1 5 7 8 W x 5 3 8 D x 5 H 8 3 8 W x 7 7 8 D x 9 H 18050 1 15 x 14 x 13 21 x 20 x 23 18100 22 0 11 5 8 W x 19 1 2...

Страница 8: ...quipment for the heating of materials per the provisions of the Electromagnetic Compatibility Directive 89 336 EEC as amended by 92 31 EEC and 93 68 EEC and per the provisions of the Low Voltage Direc...

Страница 9: ...nform with local code and configuration require ments If a plug must be installed use only the 3 prong grounded type rated for the unit load requirements and matching the power outlet Make sure the gr...

Страница 10: ...causing death or severe injury Warning Do not heat any substance above a temperature that will cause it to emit toxic fumes Death or severe injury may result Warning At higher temperatures the exterio...

Страница 11: ...e water at the moment The new setpoint is established Temperature Calibration Digital Water Baths Place a thermometer in the approximate geomet ric center of the water pan allow at least 1 full hour f...

Страница 12: ...Temperature is Regulated by Two Thermostats The control panel consists of two thermostats control and high limit and two pilot lights that indicate which thermo stat is controlling temperature The di...

Страница 13: ...limit thermo stat from its full clockwise direction slowly coun terclockwise until the point where the amber sta tus lamp goes out and the red hi limit status lamp goes ON Rotate the high limit thermo...

Страница 14: ...well Dry thoroughly The Alloy Called Stainless Stainless steel is an alloy of steel with chromium and nickel that increase the metal s resistance to rust and cor rosion Yet if not properly cared for...

Страница 15: ...ease pH to a 7 to 9 value Sodium carbonate or sodium bicarbonate may be used but they tend to form scale that must be rinsed out regularly If pH is 10 0 add sodium bisulfate to decrease pH to a 7 to 9...

Страница 16: ...rs for scale and rust removal Allow solution to soak the surface affected until rust and scale is loosened Immediately follow with a clean water rinse Citric acid based cleaners Bathroom tub and tile...

Страница 17: ...s of the approach utilized ALWAYS follow the manufacturer s directions and allow the chemicals to do the cleaning with MINIMAL scrubbing Always follow cleanings with a clean water rinse Air dry Materi...

Страница 18: ...ed hi limit and the user adjustable hi limit thermostats fail there is a third over temperature thermostat The thermostat is located underneath the water bath s bottom panel attached to the pan If the...

Страница 19: ...18052 720 433 01 Over temperature Protection OTP 18000 18000 1 18005 18005 1 18010 18010 1 18100 18100 1 18800 18800 1 18900 18900 1 18002 18002 1CE 18007 18007 1CE 18012 18012 1CE 18102 18102 1CE 182...

Страница 20: ...790 225 00 Status Lamp Base 120V 240V 360 233 01 Status Lamp Lens Red 360 234 00 Status Lamp Temperature Amber 360 235 00 Thermometer Grommet 790 247 00 Switch Power 120V Units 440 359 00 Switch Powe...

Страница 21: ...rom whom you pur chased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle you...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...sentative ii misuse including use inconsistent with written operating instructions for the product mishan dling contamination overheating modification or alteration of the product by any customer or t...