background image

1

Type S108500 SAFE-T S10

Explosion Proof Stirrer

LT1085X1 • 4/27/05

Serial Number  _____________________________________

OPERATION MANUAL

SERIES 1085

UL Listed for Class I,

Group C and D Hazardous Locations

Model #

Voltage

Description

S108525

100-120

10.5” X 10.5” (26.7 cm X 26.7 cm)

UL Listed Ceramic Top

Содержание 1085 Series

Страница 1: ...r LT1085X1 4 27 05 Serial Number _____________________________________ OPERATION MANUAL SERIES 1085 UL Listed for Class I Group C and D Hazardous Locations Model Voltage Description S108525 100 120 10 5 X 10 5 26 7 cm X 26 7 cm UL Listed Ceramic Top ...

Страница 2: ... General Usage 5 Principles of Operation 5 General Specifications 6 Installation 7 Site Selection 7 Operation 8 Stirring Speed Specifications 8 Setting the Stirring Speed 8 Stirring Speed Guide 8 Overcurrent Protection 9 Maintenance and Servicing 10 General Cleaning Instructions 10 Ordering Procedures 11 One Year Limited Warranty 12 ...

Страница 3: ... with installing Explosion Proof Flameproof electrical equipment and insures compatibility among stirrer specifications power source and ground requirements To avoid personal injury 1 Caution Due to the nature of this product its safety will be compromised should maintenance or servicing be performed on it If maintenance or servicing is need ed the stirrer must be returned to the fac tory 2 Cautio...

Страница 4: ...tone 465 C 869 F D Acrylonitrile 481 C 898 F D Ammonia 498 C 928 F D Benzene 498 C 928 F D Butane 288 C 550 F D 1 Butanol 343 C 650 F D 2 Butanol 405 C 761 F D n Butyl Acetate 421 C 790 F D Cyclopropane 503 C 938 F D Ethane 472 C 882 F D Ethanol 363 C 685 F D Ethyl Acetate 427 C 800 F D Ethylene Dichloride 413 C 775 F D Gasoline 280 471 C 536 880 F D Heptane 204 C 399 F D Hexane 225 C 437 F D Isoa...

Страница 5: ... provide excellent coupling with a stir bar All control and electrical connections are contained within the aluminum explosion proof enclosure Introduction Caution When removing the SAFE T S10 Stirrer from the shipping carton and packaging or moving the unit to the site of installation do not lift the unit by the ceramic top plate Lifting the unit by the top plat may cause dam age to the unit Note...

Страница 6: ... Electrical Ratings Volts 100 120 Amps 6 Watts 70 Phase 1 Frequency 60 Stirring Speed rpm 60 1200 Environmental Conditions Operating 17 C to 27 C 63 F to 81 F 20 to 80 relative humidity non condensing Installation Category II over voltage in accordance with IEC 664 Pollution Degree 2 in accordance with IEC 664 Altitude limit 2 000 meters Main Supply Voltage Maximum Fluctuation 10 Storage 25 C to 6...

Страница 7: ...ply circuit Overcurrent protection devices should be included in the installation The unit can be con nected to a branch circuit not exceeding 20 Amps The stirrer installation must conform with installa tion requirements for electrical equipment intend ed for use in Class I Group C and D hazardous locations as defined by the NEC and electrical code requirements of your area Installation Caution To...

Страница 8: ... stirring speed measurements made at 21 5 C Operation Maximum speed obtained without the stir bar losing coupling with the magnet in the SAFE T S10 Unit Flask Stir Bar Amount of Viscosity Maximum S108525 1 liter 2 600 ml water 1 200 rpm S108525 1 liter 2 600 ml corn oil 1 200 rpm S108525 1 liter 2 600 ml 30 wt motor oil 950 rpm S108525 500 ml 2 350 ml 1 265 centipose 400 rpm viscosity std S108525 ...

Страница 9: ...rcuit breaker back on If the overcurrent protection device is repeatedly activated during normal op erating conditions return the unit to the factory Warning To prevent ignition of hazardous atmos pheres the aluminum explosion proof enclosure chambers must be kept closed screws securely tightened whenever the unit is in operation and or connected to power supply circuit Warning To prevent ignition...

Страница 10: ...cing is needed the stirrer must be returned to the factory Disconnect the unit from the supply circuit and any installation mountings and call Barnstead International cus tomer service 1 800 553 0039 or 1 563 556 2241 General Cleaning Instructions Wipe exterior surfaces with a slightly dampened cloth containing a mild soap solution Caution To reduce the risk of the ignition any repairs made to thi...

Страница 11: ...from whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request then contact our Customer Service Department at 563 556 2241 or 800 553 0039 Prior to returning any materials to Barnstead International please contact our Customer Service Department for a ...

Страница 12: ...e ii misuse including use inconsistent with written operating instructions for the product mishandling contamination overheating modification or alteration of the product by any customer or third party or iii use of replacement parts that are obtained from a party who is not an authorized dealer of BARNSTEAD Heating elements because of their susceptibility to overheating and contamination must be ...

Отзывы: