Service work
30
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: [email protected]
Service manual
Barkey warming center III DTM
7.6
Replacing the mechanical overtemperature protection
Removal:
►
Remove the heater insert (see Chapter 7.2).
►
Remove the cover plate (see Chapter 7.2).
►
Loosen the four heater unit screws
and pull the heater unit out towards
the front.
Figure 29: Heater unit fixing screws
►
Loosen the screw which fixes the
overtemperature protection to the
heater unit.
Figure 30: Overtemperature protection fixing screw
►
Remove the mechanical overtemperature protection from the heater unit.
►
Pull both connecting cables off the overtemperature protection.
►
Unscrew the cylindrical metal adapter from the overtemperature protection.
Refitting the overtemperature protection:
Refitting the overtemperature protection takes place in reverse order to removal:
►
Screw the cylindrical metal adapter to the overtemperature protection.
Содержание warming center III DTM
Страница 2: ......