BARDOMED Care Pump EXPERT8 Скачать руководство пользователя страница 1

Lymph Drainage Therapy Device 

(Pressotherapy Device) 

User Manual

UM_EN_CPEXPERT8_01

Содержание Care Pump EXPERT8

Страница 1: ...Lymph Drainage Therapy Device Pressotherapy Device User Manual UM_EN_CPEXPERT8_01...

Страница 2: ...vice configuration 13 4 1 Main screen 13 4 2 Work mode setting with provided programs 13 4 2 2 Provided programs 13 4 2 3 User programs 14 4 2 4 Favorite programs 14 4 2 5 Create a program 14 4 3 Pres...

Страница 3: ...time speed interval hold 33 4 13 3 Chamber selection 34 5 Safety warnings 35 5 1 Energy security 35 5 2 Possibility of injuries 36 5 3 Safe use and maintenance 37 5 4 Correct use of the cuffs 37 6 Tec...

Страница 4: ...tients athletes and all other users CarePump devices are used for regeneration relaxation and improving your well being The device works very similar to manual massage which can be performed on your o...

Страница 5: ...with the cuffs Cuffs used on individual parts of the body arms legs etc By sequential inflation of the cuff chambers a part of the body where the cuff is placed is being compressed to allow achieving...

Страница 6: ...uses a sequential pressure effect which turns into a rhythmic wave Lymphatic drainage massage is pos itively perceived as a very pleasant and effective way to re generate and relax The patient feels r...

Страница 7: ...ebitis skin diseases moles lymphangitis dermatitis wounds arthritis high body temperature thick external visible varicose veins cardiac rhythm disturbances arrhythmia blood pressure disorders myocardi...

Страница 8: ...l be fore starting work with the device 2 1 CarePump Expert8 device front view main power switch input to connect the air ducts control keyboard touch screen display 2 2 CarePump Expert8 device back v...

Страница 9: ...dequate for its weight Remember not to place the device too close to the wall because the holes of cooling fans are located on the back of the case Do not place on or near the device other electrical...

Страница 10: ...er end of the cord into a power socket Attention Use the supplied power cord or one that complies with electrical standards Make sure that the voltage in the installation is 100 240V 50 60Hz If the ca...

Страница 11: ...ing the air duct or the splitter pay attention to the notch which indicates proper connection of them To connect the air duct push it into the correct input until you hear a click of both side button...

Страница 12: ...f the cuff with correct order The proper con nection is indicated by a dark gray plug The separate ends should be connected in following order from the longest to the shortest one Correctly plugged en...

Страница 13: ...hen the main screen view will appear where you can set basic treatment options 4 2 2 Provided programs 4 2 2 1 Setting using the keyboard Press any key on the keyboard to activate the icon named Progr...

Страница 14: ...Create new program 4 2 2 2 Setting using the touch screen Press the icon Programs in the upper left corner of the display Then following screen will appear where you can select provided programs user...

Страница 15: ...OK button Attention Pressure as a parameter is not defined for provided pro grams It should be adapted to the needs of each patient based on the patient s blood pressure and his feelings dur ing the...

Страница 16: ...on the icon a scrolling list with values 5 105 minutes will be displayed on the screen Values on the list can be displayed with an ac curacy of 1 5 and 10 You can decide on the accuracy of the display...

Страница 17: ...ner of the display and then select one of the body areas highlighted in white B When you press arm leg hip and abdomen area or hip and buttock area the selected body part will be highlight ed in blue...

Страница 18: ...amber press the corresponding area on the graphic image showing the cuff A scrolling list with pressure values in mmHg will be displayed on the screen Once the required value of the pressure is highli...

Страница 19: ...will drop gradually starting from the value set in starting pressure up to the value specified in fi nal pressure The pressure drop can be set in the pressure unit mmHg or as a percentage value It can...

Страница 20: ...3 4 5 6 7 8 7 8 10 9 11 OK 4 9 Interval Hold function The parameter Interval A allows you to precisely de fine the time between subsequent cycles of filling the chambers The parameter Hold B allows yo...

Страница 21: ...e language press lan guage and select the desired option from the list The ac tive language will be marked with symbol SETTINGS v2 3 0 Language ENG Sounds Password management Other CHANGE LANGUAGE Pol...

Страница 22: ...confirm it by pressing OK button Device is ready turning on off the device standby sound after start up and selecting the volume of it End of therapy turning on off the acoustic sig nals after the en...

Страница 23: ...click on the icon with crossed eye next to the asterisks symbolizing the pass word The manufacturer recommends saving the password in a safe place inaccessible for unauthorized persons If you set a p...

Страница 24: ...tion that allows you to select whether after starting the device the home screen will be the main screen or a list of available programs Keyboard brightness function that allows you to set the keyboar...

Страница 25: ...n B The treatment will stop and the treatment time will be paused The treatment can be resumed by pressing the icon again To permanently stop the treatment press the icon C When treatment is completel...

Страница 26: ...I The course of the treatment Single filling of consecutive chambers while keeping the pressure in previously inflated chambers Time 50 minutes Speed 5 Interval 3 sec Hold 3 sec Gradient 1 mmHg 2 LYM...

Страница 27: ...chamber keeps the pressure Time 50 minutes Speed 5 Interval 0 sec Hold 3 sec Gradient 1 mmHg 4 LYMPHOEDEMA IV The course of the treatment Double ascending wave The chambers inflate one at a time While...

Страница 28: ...previously inflated chambers Time 45 minutes Speed 5 Interval 3 sec Hold 2 sec Gradient 3 mmHg 6 VENOUS OEDEMA II The course of the treatment Simultaneous filling of two consecutive chambers while ke...

Страница 29: ...of the first inflated chamber which deflates during pumping of the third one Time 60 minutes Speed 5 Interval 3 sec Hold 3 sec Gradient 2 mmHg 8 LIPOEDEMA II The course of the treatment Simultaneous...

Страница 30: ...radient 4 mmHg 10 CELLULITE ADVANCED The course of the treatment The treatment is divided into three sections during which 2 4 6 chambers keep the pressure and the next two cham bers are filled and de...

Страница 31: ...es Speed 5 Interval 2 sec Hold 2 sec Gradient 4 mmHg 12 PREVENTION RELAXATION The course of the treatment Single ascending wave The chambers inflate one at a time As the third chamber is pumping the s...

Страница 32: ...13 ALL CHAMBER MASSAGE The course of the treatment All chambers inflate simultaneously and keep the pres sure performing a global massage Time 30 minutes Speed 5 Interval 3 sec Hold 2 sec Gradient 0 m...

Страница 33: ...eyboard dis played on the screen Confirm the name by pressing OK button 4 13 2 Setting parameters time speed interval hold In the next step set the parameters for the new program such as time speed ho...

Страница 34: ...marked with a white dot A To return to the view of the previous sequence use button 25 sequences can be set for each new program The program can be added to the list of favorite programs during creat...

Страница 35: ...network are consistent with the data provided on the rating plate at the bottom of the case The manufacturer recommends connecting the de vices to an installation equipped with a residual current dev...

Страница 36: ...e adapt ed to adults Using them on younger patients may lead to permanent injuries with an implanted artificial heart and other heart prostheses with metal or joint implants as the procedure may cause...

Страница 37: ...5 m the floor should be made of non conductive materials the ground on which the device is located should be stable made of non conductive and non slip materi als the device should not be exposed to...

Страница 38: ...Hold function 1 50 Interval function 1 50 Precise regulation of pressure and time in steps of 1 5 or 10 Setting of individual pressure in each chamber included Number of independent channels 1 Adjust...

Страница 39: ...39 The basic Carepump set includes Central unit device Power cord 7 Contents of the package A B A B...

Страница 40: ...th air duct hip and abdomen cuff with air duct short pants cuff with air duct safety button with a cable for the patient allowing him to stop the treatment at any time air duct splitter for simultaneo...

Страница 41: ...ing it in a manner inconsistent with the manual may result in the loss of warranty and in case of the need for intervention by an authorized service may result in treating it as a com mercial service...

Страница 42: ...her accessories contact an author ized service center or the manufacturer The touch screen works incorrectly or it does not work at all Disconnect the device from the power supply Clean the front pane...

Страница 43: ...wer input on the back of the case Below the socket for the power cord there is a fuse drawer A Open it and pull out the glass fuses placed inside If the plate inside the glass fuse is intact it means...

Страница 44: ...referably every 12 months If you experience any failure with your machine or its ac cessories or if you notice a device malfunction contact the distributor immediately In case of any disturbance in th...

Страница 45: ...e carried out by an au thorized service center with personnel trained by the man ufacturer In case of problems with locating the service please contact the device distributor Only an authorized servic...

Страница 46: ...gation as indicated by the following symbol on the label of the device The product should be disposed of in dedicated places and points for electronic waste and in accordance with the pre vailing law...

Страница 47: ...itions are met Attention 1 Before sending the machine to service contact the dis tributor and remember to include the technical passport of the device inside the package 2 Pack the device in accordanc...

Страница 48: ...Lite 4 Carepump Lite 6 are in conformity with the provisions of the following EC directives EC Electromagnetic Compatibility EMC Directive 2014 30 EU EC Low Voltage Directive LVD 2014 35 EU and that...

Страница 49: ...Manufacturer BardoMed Sp z o o ul Konecznego 6 66 31 216 Krak w NIP 945217693...