Site~Rite
®
8 Ultrasound System
Page 73
23. EMC TABLES
Warning:
The use of accessories other than those specified in the “Overview” section may result in increased
emissions or decreased immunity of the Site~Rite
®
8 Ultrasound System.
Site~Rite® 8 Ultrasound System EMC Tables
Guidance and Manufacturer’s Declaration - Emissions
The Site~Rite
®
8 Ultrasound System is intended for use in the electromagnetic environment specified below. The customer or
user of the Site~Rite
®
8 Ultrasound System should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11 / EN55011
Group 1
The Site~Rite
®
8 Ultrasound System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11 / EN55011
Class A
The Site~Rite
®
8 Ultrasound System is suitable for use in all establishments, other
than domestic, and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonics
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Immunity
The Site~Rite
®
8 Ultrasound System is intended for use in the electromagnetic environment specified below. The customer or
user of the Site~Rite
®
8 Ultrasound System should ensure that it is used in such an environment.
Immunity Test
EN/IEC 60601 Test Level
Compliance Level
Electromagnetic Environment –
Guidance
ESD
EN/IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete
or ceramic tile. If floors are
synthetic, the relative humidity
should be at least 30%.
EFT
EN/IEC 61000-4-4
±2kV Mains
±1kV Input/Output Lines
±2kV Mains
±1kV Input/Output Lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
EN/IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips/Dropout
EN/IEC 61000-4-11
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Site~Rite
®
8 Ultrasound
System requires continued
operation during power mains
interruptions, it is recommended
that the Site~Rite
®
8 Ultrasound
System be powered from an
uninterruptible power supply or
battery.
Power Frequency
50/60 Hz
Magnetic Field
EN/IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be that of a typical
commercial or hospital
environment.