17
R5914634 /01 MDSC-8527
Protective earth
Earth the MDSC-8527 by connecting the protective earth pin to a grounded outlet by means of a yellow/green
AWG18 wire (maximum admitted cable length according to national regulation requirements).
Image 2–5
CAUTION:
The display must be earthed.
Potential equalization
When potential equalization between the display and other devices is required then connect the potential
equalization pin (POAG) to the potential equalization terminal of the equipment.
Image 2–6
2.4 Cable routing
To route the cables
Route all cables through the cable routing cutout while reinstalling the connector compartment.
Image 2–7
CAUTION:
When the display is assembled in the medical system, take care of the fixation of all
cables, to avoid unwanted detachment.