6.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Type of protection (electrical):
Display with external power supply: Class I equipment.
Degree of safety (flammable anesthetic mixture)
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Non-patient care equipment
•
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
•
The equipment shall not be used with life support equipment.
•
The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the
patient at the same time.
Power connection
–
Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC (
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-
voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
High magnetic environment
•
The device shall not be used in the high magnetic environment of an MRI scanner.
•
The installer shall assess the magnetic environment before installation or use of the device.
Power cords:
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
Содержание MDNC-6121
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 6121 User Guide 5 8MP high bright color display Nio Color 5MP...
Страница 6: ...K5903146 03 Nio Color 5MP 6 Welcome...
Страница 16: ...K5903146 03 Nio Color 5MP 16 Installation...
Страница 17: ...17 K5903146 03 Nio Color 5MP Daily operation 3...
Страница 21: ...21 K5903146 03 Nio Color 5MP Advanced operation 4...
Страница 34: ...K5903146 03 Nio Color 5MP 34 Advanced operation...
Страница 35: ...35 K5903146 03 Nio Color 5MP Cleaning your display 5...
Страница 37: ...37 K5903146 03 Nio Color 5MP Important information 6...