R5906848 /04
DP2K SLP Series
320
Situation
Solution
Cooler module isn't cooling
properly.
•
Check the cooling liquid reservoir. Fill it up if necessary.
•
If error persists, replace the Light Source. See
, page 59.
Malfunction anode fan or
cathode fan.
Check the speed and voltage of the anode and cathode fan. Replace any
malfunctioning fan.
Code 5314:
“
LLS ambient - temperature sensor open
”
(Error)
Situation
Solution
Connector of the temperature
sensor is disconnected
Plug the connector of the temperature sensor into its socket on the Signal
Backplane.
Disconnected wiring between
Signal Backplane and
temperature sensor in the air
outlet channel of the projector.
Check the connection between the wire units. The connection is made
behind the back cover.
Damaged wiring of the
temperature sensor.
1.
Repair the wiring.
2.
If not repairable, replace the whole wire unit and temperature sensor.
Defect temperature sensor.
Replace the temperature sensor. Temperature sensor is visible through
the air outlet on top of the projector.
Code 5315:
“
LLS ambient - temperature sensor short
”
(Error)
Situation
Solution
Damaged insulation of the
wiring of the temperature
sensor which measures the
temperature in the channel of
the air outlet.
1.
Repair the insulation of the wire unit using shrink sleeve.
2.
If not repairable, replace the whole wire unit and temperature sensor.
Defect temperature sensor,
which measures the
temperature in the channel of
the air outlet.
Replace the temperature sensor.
Ignore any
“
temperature sensor open
”
errors that are triggered when disconnecting the temperature
sensor connector from the signal backplane.
Code 5320:
“
LCB - force light off
”
Situation
Solution
The Fan Control & Motor
board forces to switch off the
Laser Power Module due to an
Error.
Look for other errors in the log files and try to solve them first.
This can be due to:
•
Temperature too high
•
light unit not connected.
•
Light processor is not connected.
Malfunction Cinema Control
board.
Replace the Cinema Control board. See service manual chapter
Replacement of the Cinema Controller
Code 5340:
“
dmd red - temperature too high
”
(Error)
This error code is probably preceded by the warning code 5341:
“
dmd red - temperature high
”
. The same
troubleshooting table can be applied to.
Содержание DP2K SLP Series
Страница 1: ...ENABLING BRIGHT OUTCOMES Service Manual DP2K SLP Series ...
Страница 4: ......
Страница 10: ...R5906848 04 DP2K SLP Series 10 ...
Страница 14: ...R5906848 04 DP2K SLP Series 14 Safety ...
Страница 24: ...R5906848 04 DP2K SLP Series 24 General ...
Страница 30: ...R5906848 04 DP2K SLP Series 30 Preventative maintenance actions ...
Страница 38: ...R5906848 04 DP2K SLP Series 38 Mains Input ...
Страница 90: ...R5906848 04 DP2K SLP Series 90 Laser Phosphor Light Source ...
Страница 94: ...R5906848 04 DP2K SLP Series 94 10 Reinstall all covers Liquid cooling circuit ...
Страница 138: ...R5906848 04 DP2K SLP Series 138 Light Pipe ...
Страница 147: ...147 R5906848 04 DP2K SLP Series Optical alignment 12 ...
Страница 150: ...R5906848 04 DP2K SLP Series 150 Optical alignment ...
Страница 248: ...R5906848 04 DP2K SLP Series 248 Card Cage ...
Страница 263: ...263 R5906848 04 DP2K SLP Series Exhaust system 20 ...
Страница 265: ...265 R5906848 04 DP2K SLP Series Board Diagnostic LED s 21 ...
Страница 270: ...R5906848 04 DP2K SLP Series 270 Board Diagnostic LED s ...
Страница 290: ...R5906848 04 DP2K SLP Series 290 Dust Filters ...
Страница 294: ...R5906848 04 DP2K SLP Series 294 Troubleshooting ...
Страница 382: ...R5906848 04 DP2K SLP Series 382 Troubleshooting ...
Страница 394: ...R5906848 04 DP2K SLP Series 394 Projector cleaning ...
Страница 395: ...395 Overview Hazards R5906848 04 DP2K SLP Series Appendix A ...
Страница 409: ...R5906848 04 DP2K SLP Series 409 ...