18. Lamp power supply (single module)
18.1 LPS module diagnostic LED’s
Status LED’s on the LPS module
There are 8 diagnostic LED’s per LPS module. Four orange, three green and one red LED.
The orange LED “LVPS OK” lights up immediately after the projector is switched on. At the same time, the heartbeat LED starts
blinking. All other status LEDs of the LPS unit remain off. This is the standby status of the LPS unit. Once the command is sent
to the LPS units to start up the projection lamp, the green LEDs are lighting up one after the other. First the green LED “PFC OK”,
then the green LED “LPS OK” and
fi
nally, when the lamp is ignited, the green LED “LAMP ON”. The right orange of the upper row
blinks. This is the heartbeat signal
The red LED “ERR” remains off unless an error is detected inside the LPS unit or when both LPS units are ordered to shutdown due
to a malfunction somewhere else inside the projector.
- LAMP
ERR
N
STATUS
CTLB OUT
ADDRESS CTRL
IN
LAMP ON / LVPS OK
CTRL OUT
CTLB IN
PFC
OK
LPS
OK
MAINS
INPUT
3
2
1
4
5
6
7
8
Image 18-1
1
Diagnostic LED “LPS HEARTBEAT” (orange).
2
(internal use only).
3
Diagnostic LED “LPS TRANS” (orange).
4
Diagnostic LED “ERROR” (red).
5
Diagnostic LED “PFC OK” (green).
6
Diagnostic LED “LPS OK” (green).
7
Diagnostic LED “LAMP ON” (green).
8
Diagnostic LED “LVPS OK” (yellow).
Diagnostic
About the orange LEDs next to the CTRL connectors:
LED 2 is only for internal use. LED 1 is the heartbeat LED.
LPS
HEARTBEAT
(Orange LED 1)
Diagnostics
Action
Blinking
Normal operation
-
OFF
12 V from backplane via CTRL IN not available on LPS
unit
Check 12V out on backplane.
ON
12 V from backplane available.
Replace the LPS module.
LPS TRANS
(Orange LED 3)
Diagnostics
Action
Blinking
Normal operation, blinks each time a valid transmission
has been received
-
Communication fault (no reception of valid commands
from primary side)
Replace LPS unit
Fault on primary side of LPS (E.g. fuse blown and no
mains voltage available inside LPS)
Replace fuse.
If the problem is not solved, replace LPS
OFF or ON
No mains voltage present at input of LPS mains
terminals
Check main voltage.
About the diagnostic LEDs, ERR, PFC, LPS, Lamp ON and LVPS.
292
R5905043 DP2K-12C/11CX 19/02/2018
Содержание DP2K-12C
Страница 1: ...DP2K 12C 11CX Service manual R5905043 15 19 02 2018...
Страница 4: ......
Страница 8: ...Table of contents 4 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 12: ...1 Safety 8 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 31: ...4 Troubleshooting 4 TROUBLESHOOTING R5905043 DP2K 12C 11CX 19 02 2018 27...
Страница 32: ...4 Troubleshooting 28 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 96: ...5 Removal and installation of projector covers 92 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 121: ...6 Lamps and lamp houses Image 6 38 R5905043 DP2K 12C 11CX 19 02 2018 117...
Страница 138: ...7 Cold mirror assembly 134 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 175: ...10 Integrator rod 1 2 3 4 Image 10 4 Light pipe housing 4 Take off the housing R5905043 DP2K 12C 11CX 19 02 2018 171...
Страница 222: ...12 Card Cage 1 Image 12 11 218 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 225: ...12 Card Cage Image 12 15 Button unit removal R5905043 DP2K 12C 11CX 19 02 2018 221...
Страница 281: ...15 Power input Image 15 13 Mains AC compartment cover R5905043 DP2K 12C 11CX 19 02 2018 277...
Страница 282: ...15 Power input 278 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 299: ...18 Lamp power supply single module Image 18 4 LPS removal R5905043 DP2K 12C 11CX 19 02 2018 295...
Страница 306: ...18 Lamp power supply single module 302 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 340: ...21 Maintenance procedures 336 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 350: ...A Pin configurations 346 R5905043 DP2K 12C 11CX 19 02 2018...
Страница 351: ...B Input formats B INPUT FORMATS R5905043 DP2K 12C 11CX 19 02 2018 347...
Страница 352: ...B Input formats 348 R5905043 DP2K 12C 11CX 19 02 2018...