8.1 Safety information
General information
This device is for use in professional healthcare facility environments only.
With the installation of the device, use only the delivered external cables and power supply or a spare part
provided by the legal manufacturer. Using another can result in a decrease of the immunity level of the device.
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock, Fire Hazard and Mechanical Hazard
Only in accordance with ANSI/AAMI ES 60601-1 (2005) + AMD (2012) and CAN/CSA-C22.2 No.
60601-1: 14 (Label UL Mark)
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Make sure the electrical contacts of the grip / battery pack and battery charging station are dry and clean to
avoid short circuit.
Keep the On/Off button on top of the head pressed to safely shut down the Demetra Scope.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Inspection and maintenance
Inspect the display and contact glass of the Demetra Scope for damage or scratches at regular intervals.
Properly clean and disinfect the Demetra Scope. See
”
, page 47
for more details.
Type of protection (electrical):
The Demetra Scope has an insulation protection class II power supply adapter.
Power connection
•
Power requirements: The equipment must be powered using the delivered adapter and appropriate plug
(for your country).
•
The adapter must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
•
Disconnecting the adapter from the building mains outlet is the proper means of isolation from the mains
supply.
Batteries and charging
Your Demetra Scope uses lithium batteries (one in the grip and one in the head), which can only be purchased
from Barco or from an authorized Barco dealer. Do not under any circumstances use other batteries than the
ones designed for this unit. Doing so could result in damage to the unit.
Do not place your Demetra Scope in or near a heat source. Excessive heating can damage the device and/or
the batteries inside the device. Exposing these components to excessive heat for a prolonged period could
Содержание Demetra
Страница 1: ...ENABLING BRIGHT OUTCOMES BDEM 01 User Guide Barco Demetra...
Страница 10: ...K5902200 07 Barco Demetra 10 Welcome...
Страница 42: ...K5902200 07 Barco Demetra 42 Troubleshooting...
Страница 43: ...43 K5902200 07 Barco Demetra 6 1 Repacking overview 44 Repacking instructions 6...
Страница 45: ...45 K5902200 07 Barco Demetra 11 Plugs Repacking instructions...
Страница 46: ...K5902200 07 Barco Demetra 46 Repacking instructions...
Страница 47: ...47 K5902200 07 Barco Demetra 7 1 Instructions 48 Cleaning and disinfection 7...
Страница 66: ...K5902200 07 Barco Demetra 66 Important information...
Страница 67: ......
Страница 68: ...www barco com K5902200 07 2021 11 04 Barco NV President Kennedypark 35 8500 Kortrijk Belgium K5902200 07...