Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
4.18.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Compliance Check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
4.18.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM calibration
to start the calibration.
Warning:
Pressing a key during calibration will abort the calibration, previous values will be restored.
Note:
After calibration, the compliance check will start automatically.
Содержание Coronis Fusion 4MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 MDCC 6530 User guide Coronis Fusion 4MP 6MP...
Страница 5: ...5 R5911227 02 Coronis Fusion 4MP 6MP Welcome 1...
Страница 8: ...R5911227 02 Coronis Fusion 4MP 6MP 8 Welcome...
Страница 16: ...R5911227 02 Coronis Fusion 4MP 6MP 16 Installation...
Страница 17: ...17 R5911227 02 Coronis Fusion 4MP 6MP Daily operation 3...
Страница 21: ...21 R5911227 02 Coronis Fusion 4MP 6MP Advanced operation 4...
Страница 34: ...R5911227 02 Coronis Fusion 4MP 6MP 34 Advanced operation...
Страница 35: ...35 R5911227 02 Coronis Fusion 4MP 6MP Cleaning your display 5...
Страница 37: ...37 R5911227 02 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Страница 41: ...R5911227 02 Coronis Fusion 4MP 6MP 41 Image 6 7 Repackaging instructions...
Страница 42: ...R5911227 02 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Страница 43: ...43 R5911227 02 Coronis Fusion 4MP 6MP Important information 7...
Страница 59: ...R5911227 02 Coronis Fusion 4MP 6MP 59 Important information...