TDSscan20 Pocket TDS Tester
USER MANUAL
Bante Instruments Inc.
Страница 1: ...TDSscan20 Pocket TDS Tester USER MANUAL Bante Instruments Inc...
Страница 2: ...u Select an option Confirm the calibration settings or displayed option Display Icon Description When the battery voltage falls below the minimum power requirements the icon automatically disappears M...
Страница 3: ...the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode press and hold the Cal key to enter the setup menu 2 If necessary press the Cal key again to...
Страница 4: ...ation is completed Multipoint Calibration Ensure that you have selected 2 or 3 points calibration in the setup menu 2 1 Repeat steps 1 1 through 1 5 above When the first calibration point is completed...
Страница 5: ...l automatically lock the measurement endpoint and show HOLD icon Press the Meas key to resume measuring Electrode Maintenance and Replacement Rinse the electrode thoroughly with distilled water after...
Страница 6: ...45 g Fill the distilled water to the mark mix the solution until the reagent is completely dissolved Optional Accessories Order Code Description E ECscan C1 10K Platinum conductivity cell K 1 TDSCS 71...
Страница 7: ...ster is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your tester s...