20
english
10.1 Accuracy
The specifications are typical values for the
BTL7-A/C/E/G… at 24 V DC and room temperature, with
a nominal length of 500 mm in conjunction with the
BTL-P-1013-4R, BTL-P-1013-4S, BTL-P-1012-4R or
BTL-P-1014-2R magnet.
The BTL is fully operational immediately, with full accuracy
after warm-up.
For special versions, other technical data may
apply.
Special versions are indicated by the suffix -SA
on the part label.
Repeat accuracy
Typical
±10 µm
Sampling rate
Dependent on the nominal length
At nominal length = 500 mm
250 µs…5.7 ms
500 µs
Non-linearity at
Nominal length ≤ 500 mm
Nominal length > 500 to ≤ 5500 mm
Nominal length > 5500 mm
±50 µm
±0.01% FS
±0.02% FS
Temperature coefficient
1)
≤ 30 ppm/K
Max. detectable velocity
10 m/s
10.2 Ambient conditions
Ambient temperature
−40…+85°C
Ambient temperature for UL
(only BTL7-…-KA…)
≤ +80°C
Storage temperature
−40…+100°C
Relative humidity
≤ 90%,
non-condensing
Rod pressure rating (when installed in
hydraulic cylinders)
For Ø 8 mm
≤ 250 bar
For Ø 10.2 mm
≤ 600 bar
Shock rating
Continuous shock
per EN 60068-2-27
3), 4)
150 g/6 ms
150 g/2 ms
Vibration
per EN 60068-2-6
3), 4)
20 g,
10…2000 Hz
Degree of protection per IEC 60529
Connector SR32/SR115
(when attached)
IP67
Cable
IP68
3)
2)
10.3 Supply voltage (external)
Voltage, stabilized
5)
10…30 V DC
Ripple
≤ 0.5 V
ss
Current draw
(at 24 V DC)
< 120 mA
Inrush current
≤ 500 mA
Reverse polarity protection
6)
Up to 36 V
Overvoltage protection
Up to 36 V
Dielectric strength
(GND to housing)
500 V AC
10.4 Output
BTL7-A… Output voltage
Load current
0…10 V and 10…0 V
≤ 5 mA
BTL7-C… Output current
Load resistance
0…20 mA/20…0 mA
≤ 500 ohms
BTL7-E… Output current
Load resistance
4…20 mA/20…4 mA
≤ 500 ohms
BTL7-G… Output voltage
Load current
−10…10 V and 10…−10 V
≤ 5 mA
Short-circuit resistance
Signal cable to 36 V
Signal cable to GND
10.5 Input
Programming inputs La, Lb:
High-active
BTL7-_5_ _-…
10…30 V DC
Overvoltage protection
Up to 36 V
1)
Nominal length 500 mm, magnet in the middle of the measuring range
2)
For UL: Use in enclosed spaces and up to a height of 2000 m above
sea level.
3)
Individual specifications as per Balluff factory standard
4)
Resonant frequencies excluded
5)
For UL: The BTL must be externally connected via a limited-energy circuit
as defined in UL 61010-1, a low-power source as defined in UL 60950-1,
or a class 2 power supply as defined in UL 1310 or UL 1585.
6)
A prerequisite is that no current can flow between GND and 0 V in the
event of polarity reversal.
10
Technical data
BTL7-A/C/E/G5 _ _ -M _ _ _ _ -K(8)-SR32/SR115/_ _ _
Magnetostrictive Linear Position Sensor – Rod Style
Содержание BTL7-A510 Series
Страница 2: ...www balluff com...
Страница 3: ...Betriebsanleitung deutsch BTL7 A C E G5_ _ M_ _ _ _ K 8 SR32 SR115 _ _ _...
Страница 4: ...www balluff com...
Страница 29: ...User s Guide english BTL7 A C E G5_ _ M_ _ _ _ K 8 SR32 SR115 _ _ _...
Страница 30: ...www balluff com...
Страница 55: ...Notice d utilisation fran ais BTL7 A C E G5_ _ M_ _ _ _ K 8 SR32 SR115 _ _ _...
Страница 56: ...www balluff com...
Страница 81: ...Manuale d uso italiano BTL7 A C E G5_ _ M_ _ _ _ K 8 SR32 SR115 _ _ _...
Страница 82: ...www balluff com...
Страница 107: ...Manual de instrucciones espa ol BTL7 A C E G5_ _ M_ _ _ _ K 8 SR32 SR115 _ _ _...
Страница 108: ...www balluff com...
Страница 133: ......