Balluff BDD-UM 3023 Скачать руководство пользователя страница 8

4.2.

Hints against noisy enviroment

All in puts and out puts are pro tec ted against noisy en vi ron ment and high vol ta ge
spikes.  Ne vert he less  the  lo ca ti on  should  be  se lec ted  to  en su re  that  no  ca pa ci ti ve 
or in duc ti ve in ter fe rence can have an ef fect on the in stru ment or the con nec tion
li nes.

It is ad visab le:

• To use shiel ded cables.

• The wi ring of shields and ground (0V) should be star- sha ped.

• The di stan ce to in ter fe rence sour ces should be as long as pos si -

ble.  If  ne cessa ry,  pro tec ti ve  scre en  or  me tal  en clo su res  must  be
pro vi ded.

• Coils of re lays must be supp lied with fil ters.

• Par al lel wi ring of in put si gnals and AC po wer li nes should be

avoi ded.

4.3.

Connection and pin assignment

All in puts and out puts are con nec tors de signed as plug- in screw ter mi nals.

7

4. Electrical connections

1 2 3

4

5

6

Содержание BDD-UM 3023

Страница 1: ...english User s Guide BDD UM 3023...

Страница 2: ...efective material and workmanship for a period of two 2 years from date of delivery If it is necessary to return the product to us the sender is responsible for shipping charges freight insurance and...

Страница 3: ...ns 6 4 1 General Instructions 6 4 2 Hints against noisy enviroment 7 4 3 Connection and pin assignment 7 4 4 Connection of input signals 8 4 4 1 Input voltage 0 10 V 8 4 4 2 Input current 0 20 mA 4 20...

Страница 4: ...7 1 1 Scaling the display range 12 7 2 Tare function 13 7 3 Display test 13 7 4 Programming quick reference 13 8 Technical specifications 14 3...

Страница 5: ...lity to customer or any other person or entity with respect to any liablity loss or damage caused or alleged to be caused directly or indirectly by equipment or software sold or furnished by us Read t...

Страница 6: ...ion 3 Mounting 3 1 Place of operation Attention must be payed to the protection against humidity dust and high tempe ratures at the place of operation 3 2 Panel mounting 3 2 1 Panel for switch board I...

Страница 7: ...tages applied Cords must be provided with sleeves Attention must be paid that the power supply voltage applied will agree with voltage noticed at the Type plate It is important to ensure a careful gro...

Страница 8: ...advisable To use shielded cables The wiring of shields and ground 0V should be star shaped The distance to interference sources should be as long as possi ble If necessary protective screen or metal e...

Страница 9: ...3 Signal ground 6 Supply voltage 4 4 Connection of input signals 4 4 1 Input voltage 0 10 V 4 4 2 Input current 0 20 mA 4 20 mA 4 5 Connection of power supply voltage 8 4 Electrical connections 3 0 V...

Страница 10: ...onfiguration by programming Attention When the instrument is built in a machine and the custo mer wants to change the configuration attention must be paid that no damage will occur to the machine 6 Pr...

Страница 11: ...y the buttons on the front Push button Function P button Selection of parameter button Increase of number of parameter parameter button Decrease of number of parameter parameter button select the disp...

Страница 12: ...on Confirm the parameter by pressing the P button The display shows the programming level and the number of the parameter e g 0 04 parameter 4 7 1 Programming level for configuration 0 00 Param Descri...

Страница 13: ...input current 1mA or 3mA the display range exceeds the possible minimum 999 the parameter P0 03 has the value3 decimal point after the first decade X XXX in this case is the display of negative value...

Страница 14: ...0 04 had to be set to 1 Erase the tare value push the button for ca 3 seconds After this reset the untared value is dis played 7 3 Display test Activating the display test will drive all segments of...

Страница 15: ...n at negative Values Operation keyboard design front membrane with push buttons Power supply 24 V 20 DC isolated power consumption max 65 mA Case 96 x 48 x 60 mm depth 72 mm incl screw terminal protec...

Страница 16: ...41042 Phone 859 727 2200 Toll free 1 800 543 8390 Fax 859 727 4823 technicalsupport balluff com CN Service Center China Balluff Shanghai trading Co ltd Room 1006 Pujian Rd 145 Shanghai 200127 P R Chin...

Отзывы: