Document: 333D012G
5-18-2007
- 10 -
Environmental Conditions
The cabinet is designed for use in the following conditions:
•
Indoor use
•
Altitudes up to 2000 meters
•
Temperature range from 5
°
C to 40
°
C
•
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% at 40°C
•
Main Supply voltage 115V ±10%V AC
•
Transient over voltage according to Installation Category (OVERVOLTAGE CATEGORIES) II per
UL 61010-1, Annex J
•
Pollution Degree 2
Symbols and Terminology
Protective Earth:
Any terminal intended for connection to external protective conductor for protection
against electric shock in case of a fault.
General Caution:
Refer to instruction manual for information regarding personnel and environment
protection.
!