July 2021
20
Filter Age to 0 hrs via
the LCD touchscreen
which indicates that the
HEPA has been
replaced.
PASS CHG
n/a
This event shall be
generated if the
passcode is changed by
the user.
PASS SET CHG
n/a
This event shall be
generated when the
passcode settings on
the “Passcode
Configuration” screen,
are changed.
DATE/TIME
n/a
This indicates that the
user has changed the
date and time stored
within the RTC.
LOG DATE
<date>
Append the new date.
This event shall be
generated when the
Data Logging starting
export date is changed.
CONFIG UPD
<bits>
0-7
8
9
10
11
12-31
The argument is a 8-digit hexadecimal
value; when converted to binary, the bits
correspond to the following:
Define the
model
:
Unconfigured
= 0b11111111,
Nebulizer
= 0b00000001,
Evaporator
=
0b00000010
UV Option:
Enabled = 0b1, Disabled = 0b0
H2O2 Decon Option:
Enabled = 0b1, Disabled = 0b0
Auto Water Refill Option:
Enabled = 0b1, Disabled = 0b0
CO2 Auto SwitchOver Option
: Enabled =
0b1, Disabled = 0b0
Unused:
should be ignored.
The System
Configurable Options
have been changed.