END HEPA
n/a
The HEPA Filter Age is equal to
100% of the HEPA’s rated lifespan,
8760 hours.
PUMP ALARM
n/a
The Water Pump is in alarm state.
State/Status Events
H
2
0
2
Event Name
Event
Code
Event Code Meaning
Description
H
2
O
2
START
This event shall be generated when
the UI’s “Decon Start” (36.0
2
)
screen’s “OK” button is pressed.
This denotes that the H
2
O
2
Decontamination has started.
H
2
O
2
ABORT
This event shall be generated when
the UI’s “Decon Abort” (36.14)
screen’s “OK” button is pressed.
This denotes that the user has
aborted the H
2
O
2
Decontamination.
H
2
O
2
END
This event shall be generated when
UI’s “Decon Cycle Complete” (37.3)
screen’s “OK” button is pressed This
denotes that the H
2
O
2
Decontamination has successfully
completed.
H
2
O
2
FAILURE
0
While not in the H2O2 State
DESTRUCT_1, DESTRUCT_2,
REMOVE/COMPLETE, or
This event shall be generated to
denote that the H
2
O
2
Decontamination has failed.