PanaFlow™ LC User’s Manual
39
Chapter 4. Error Codes and Troubleshooting
S11: Clock Frequency
Clock frequency error
Fault:
Input clock frequency failure. Try power
cycling the meter. If error persists after power cycle,
contact Panametrics factory
S12: CPU
CPU error
Fault:
CPU registers have stuck bits. Try power
cycling the meter. If error persists after power cycle,
contact Panametrics factory
S13: Invariable Flash
Memory
Flash memory fault
Fault:
Flash memory test failed. Try power cycling
the meter. If error persists after power cycle, contact
Panametrics factory
S14: Invariable SRAM
Invariable SRAM fault
Fault:
Invariable SRAM memory test failed. Try power
cycling the meter. If error persists after power cycle,
contact Panametrics factory
S15: Variable Memory
Variable SRAM fault
Fault:
Variable SRAM test failed. Try power cycling
the meter. If error persists after power cycle, contact
Panametrics factory
S16: FPGA Config
FPGA configuration error
Fault:
FPGA configuration validation failure. Try
power cycling the meter. If error persists after power
cycle, contact Panametrics factory
S17: Temperature
Temperature error
Error:
Temperature of the electronics is outside the
pre-defined operating range. Make sure that the
ambient temperature is not outside the meter
operating range
S18: Driver Fault
Driver failure
Fault:
Driver failure. Try power cycling the meter. If
error persists after power cycle, contact
Panametrics factory
S19: Watch Dog Failure
Watch dog failure
Fault:
Watch dog test failed. Try power cycling the
meter. If error persists after power cycle, contact
Panametrics factory
S21: Stack Overflow
Stack overflow
Fault:
Stack overflow. Try power cycling the meter. If
error persists after power cycle, contact
Panametrics factory
S22: Sequence or Window
Watchdog
Sequence failed
Fault:
Sequence failure detected. Try power cycling
the meter. If error persists after power cycle, contact
Panametrics factory
S23: Initialization Failed
Initialization failed
Error:
Initialization failed. Please verify all the
configuration parameters. If error persists, contact
Panametrics factory
S24: DSP Hardware Errors DSP hardware failed
Fault:
DSP hardware failure detected. Try power
cycling the meter. If error persists after power cycle,
contact Panametrics factory
S25: DSP Exception
DSP exception
Fault:
DSP exception. Try power cycling the meter. If
error persists after power cycle, contact
Panametrics factory
S26: Default ISR
Exception within the ISR
Fault:
Exception within the ISR. Try power cycling the
meter. If error persists after power cycle, contact
Panametrics factory
S27: DSP Reset ISR
Exception within the DSP ISR
Fault:
Exception within the DSP ISR. Try power cycling
the meter. If error persists after power cycle, contact
Panametrics factory
S28: Software Fault
Software malfunction
Error:
Software malfunction. Try power cycling the
meter. If error persists after power cycle, contact
Panametrics factory
Table 4: System Error Description and Recommended Actions
Error Code
Error Message
Description / Recommended Action
Содержание Panametrics PanaFlow LC
Страница 1: ...panametrics com BH029C11 EN C PanaFlow LC User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 54: ...Chapter 5 Maintenance and Service 46 PanaFlow LC User s Manual no content intended for this page...
Страница 58: ...Appendix B Digital Communication 50 PanaFlow LC User s Manual no content intended for this page...
Страница 62: ...Appendix C Error code bit field representation 54 PanaFlow LC User s Manual no content intended for this page...
Страница 64: ...Appendix D CE Mark Compliance 56 PanaFlow LC User s Manual no content intended for this page...
Страница 66: ...Warranty 58 PanaFlow LZ User s Manual no content intended for this page...
Страница 67: ......