Cleaning and maintenance
7.
Cleaning the workstation
7.1
To ensure that your Invivo
2
Plus workstation remains at optimum working conditions, it must be
cleaned on a regular basis. A basic clean is required after each use. Deep cleaning is required at
regular intervals, dependent upon the nature of the materials used in the workstation. As a guide, a
deep clean should be performed at between 3-6 month intervals.
Cleaning agents
7.1.1
The correct cleaning agents must be used to clean the workstation. The use of incorrect cleaning
agents will damage the workstation and invalidate the warranty. The following cleaning agents are
permitted;
•
Ethanol, laboratory grade at a maximum concentration of 70% by volume ethanol in distilled
water.
•
Propanol, laboratory grade at a maximum concentration of 70% by volume propanol in
distilled water.
•
Tristel Fuse Sachet, 1 sachet diluted in 3 litres of distilled water, or Tristel Duo Foamer.
Tristel Fuse Sachets and Duo Foamer are available from Ruskinn Technology Limited, see
section 7.6.2 for details.
•
Ruskinn Technology Limited anti-static cleaner.
•
Distilled water.
No other cleaning agents are permitted. The use of UV light is not permitted in the workstation, as it
will damage the acrylic shell.
Cleaning procedure – during and after each use
7.1.2
During use, clean any spills immediately using paper towels soaked in an appropriate cleaning agent.
Wipe dry using a dry paper towel.
After each use;
•
Remove all waste materials from the workstation chamber.
•
Wipe the workstation chamber floor
•
Wipe the workstation chamber floor tray clean using paper towels to dry.
•
Wipe the interlock floor tray using paper towels soaked in an appropriate cleaning agent.
•
Wipe the interlock floor tray using paper towels to dry.
Note: It is easiest to clean the interlock from the outside.